Slit-scan flow cytometry has been used to measure the distribution of fluorescent dye along mammalian chromosomes. In this procedure, the chromosomes are forced to flow lengthwise and one at a time through a thin laser beam. The resulting fluorescence profile is an approximation of the distribution of stain along the chromosome showing a dip at the centromere region. In this paper we describe a procedure, based upon Fourier analysis, to restore high spatial frequency information to the measured chromosome profile so that the centromeric dip can be precisely located. To accomplish this we estimate both the flow velocity and the laser beam profile. The procedure is applied to the restoration of profiles of fluorescent microspheres and Chinese hamster M3-1 chromosomes. Validation of the procedure is accomplished by comparing restored chromosome profiles to those obtained through high resolution fluorescence microscopy.

Original languageEnglish (US)
Pages (from-to)101-106
Number of pages6
JournalIEEE Transactions on Biomedical Engineering
VolumeBME-29
Issue number2
DOIs
StatePublished - 1982
Externally publishedYes

Fingerprint

Flow cytometry
Chromosomes
Restoration
Laser beams
Fourier analysis
Fluorescence microscopy
Microspheres
Flow velocity
Dyes
Fluorescence

ASJC Scopus subject areas

  • Biomedical Engineering

Cite this

Restoration of Profiles from Slit-Scan Flow Cytometry. / Norgren, Richard M.; Gray, Joe; Young, Ian T.

In: IEEE Transactions on Biomedical Engineering, Vol. BME-29, No. 2, 1982, p. 101-106.

Research output: Contribution to journalArticle

Norgren, Richard M. ; Gray, Joe ; Young, Ian T. / Restoration of Profiles from Slit-Scan Flow Cytometry. In: IEEE Transactions on Biomedical Engineering. 1982 ; Vol. BME-29, No. 2. pp. 101-106.
@article{6c30a3a4844d4143a0d0719b3e52a4e3,
title = "Restoration of Profiles from Slit-Scan Flow Cytometry",
abstract = "Slit-scan flow cytometry has been used to measure the distribution of fluorescent dye along mammalian chromosomes. In this procedure, the chromosomes are forced to flow lengthwise and one at a time through a thin laser beam. The resulting fluorescence profile is an approximation of the distribution of stain along the chromosome showing a dip at the centromere region. In this paper we describe a procedure, based upon Fourier analysis, to restore high spatial frequency information to the measured chromosome profile so that the centromeric dip can be precisely located. To accomplish this we estimate both the flow velocity and the laser beam profile. The procedure is applied to the restoration of profiles of fluorescent microspheres and Chinese hamster M3-1 chromosomes. Validation of the procedure is accomplished by comparing restored chromosome profiles to those obtained through high resolution fluorescence microscopy.",
author = "Norgren, {Richard M.} and Joe Gray and Young, {Ian T.}",
year = "1982",
doi = "10.1109/TBME.1982.325015",
language = "English (US)",
volume = "BME-29",
pages = "101--106",
journal = "IEEE Transactions on Biomedical Engineering",
issn = "0018-9294",
publisher = "IEEE Computer Society",
number = "2",

}

TY - JOUR

T1 - Restoration of Profiles from Slit-Scan Flow Cytometry

AU - Norgren, Richard M.

AU - Gray, Joe

AU - Young, Ian T.

PY - 1982

Y1 - 1982

N2 - Slit-scan flow cytometry has been used to measure the distribution of fluorescent dye along mammalian chromosomes. In this procedure, the chromosomes are forced to flow lengthwise and one at a time through a thin laser beam. The resulting fluorescence profile is an approximation of the distribution of stain along the chromosome showing a dip at the centromere region. In this paper we describe a procedure, based upon Fourier analysis, to restore high spatial frequency information to the measured chromosome profile so that the centromeric dip can be precisely located. To accomplish this we estimate both the flow velocity and the laser beam profile. The procedure is applied to the restoration of profiles of fluorescent microspheres and Chinese hamster M3-1 chromosomes. Validation of the procedure is accomplished by comparing restored chromosome profiles to those obtained through high resolution fluorescence microscopy.

AB - Slit-scan flow cytometry has been used to measure the distribution of fluorescent dye along mammalian chromosomes. In this procedure, the chromosomes are forced to flow lengthwise and one at a time through a thin laser beam. The resulting fluorescence profile is an approximation of the distribution of stain along the chromosome showing a dip at the centromere region. In this paper we describe a procedure, based upon Fourier analysis, to restore high spatial frequency information to the measured chromosome profile so that the centromeric dip can be precisely located. To accomplish this we estimate both the flow velocity and the laser beam profile. The procedure is applied to the restoration of profiles of fluorescent microspheres and Chinese hamster M3-1 chromosomes. Validation of the procedure is accomplished by comparing restored chromosome profiles to those obtained through high resolution fluorescence microscopy.

UR - http://www.scopus.com/inward/record.url?scp=0020332080&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0020332080&partnerID=8YFLogxK

U2 - 10.1109/TBME.1982.325015

DO - 10.1109/TBME.1982.325015

M3 - Article

C2 - 7056552

AN - SCOPUS:0020332080

VL - BME-29

SP - 101

EP - 106

JO - IEEE Transactions on Biomedical Engineering

JF - IEEE Transactions on Biomedical Engineering

SN - 0018-9294

IS - 2

ER -