OtoID: New extended frequency, portable audiometer for ototoxicity monitoring

Marilyn F. Dille, Peter Jacobs, Samuel Y. Gordon, Wendy J. Helt, Garnett P. McMillan

Research output: Contribution to journalArticle

6 Citations (Scopus)

Abstract

Portability of equipment is an increasingly important component in the practice of audiology. We report on a new device, the OtoID, that supports evidence-based ototoxicity testing protocols, provides capability for hearing testing on the hospital treatment unit, and can automate patient self-testing. The purpose of this article is to report on the validation and verification of the OtoID portable audiometer in 40 subjects both young and old, with and without hearing impairment. Subjects were evaluated by an audiologist using the manual hearing test program and then self-tested via an automated testing program. Testing was done in a sound booth and on a hospital treatment unit. Therefore, data were collected in four conditions (booth vs hospital unit and automated vs manual testing) and analyzed for testing bias, repeatability, and American Speech-Language-Hearing Association-significant ototoxicity false-positive rate. Repeatable hearing threshold results were obtained on all subjects who performed the test, regardless of hearing status or testing location.

Original languageEnglish (US)
Pages (from-to)997-1006
Number of pages10
JournalJournal of Rehabilitation Research and Development
Volume50
Issue number7
DOIs
StatePublished - 2013

Fingerprint

Hospital Units
Hearing Tests
Hearing
American Speech-Language-Hearing Association
Audiology
Equipment and Supplies
Hearing Loss
Therapeutics

Keywords

  • Audiometer
  • Chemotherapy
  • Cisplatin
  • Hearing loss
  • Monitoring
  • OtoID
  • Ototoxicity
  • Rehabilitation
  • Tinnitus
  • Veterans

ASJC Scopus subject areas

  • Rehabilitation

Cite this

OtoID : New extended frequency, portable audiometer for ototoxicity monitoring. / Dille, Marilyn F.; Jacobs, Peter; Gordon, Samuel Y.; Helt, Wendy J.; McMillan, Garnett P.

In: Journal of Rehabilitation Research and Development, Vol. 50, No. 7, 2013, p. 997-1006.

Research output: Contribution to journalArticle

Dille, Marilyn F. ; Jacobs, Peter ; Gordon, Samuel Y. ; Helt, Wendy J. ; McMillan, Garnett P. / OtoID : New extended frequency, portable audiometer for ototoxicity monitoring. In: Journal of Rehabilitation Research and Development. 2013 ; Vol. 50, No. 7. pp. 997-1006.
@article{b75c301612574c279925c3ed377f5c59,
title = "OtoID: New extended frequency, portable audiometer for ototoxicity monitoring",
abstract = "Portability of equipment is an increasingly important component in the practice of audiology. We report on a new device, the OtoID, that supports evidence-based ototoxicity testing protocols, provides capability for hearing testing on the hospital treatment unit, and can automate patient self-testing. The purpose of this article is to report on the validation and verification of the OtoID portable audiometer in 40 subjects both young and old, with and without hearing impairment. Subjects were evaluated by an audiologist using the manual hearing test program and then self-tested via an automated testing program. Testing was done in a sound booth and on a hospital treatment unit. Therefore, data were collected in four conditions (booth vs hospital unit and automated vs manual testing) and analyzed for testing bias, repeatability, and American Speech-Language-Hearing Association-significant ototoxicity false-positive rate. Repeatable hearing threshold results were obtained on all subjects who performed the test, regardless of hearing status or testing location.",
keywords = "Audiometer, Chemotherapy, Cisplatin, Hearing loss, Monitoring, OtoID, Ototoxicity, Rehabilitation, Tinnitus, Veterans",
author = "Dille, {Marilyn F.} and Peter Jacobs and Gordon, {Samuel Y.} and Helt, {Wendy J.} and McMillan, {Garnett P.}",
year = "2013",
doi = "10.1682/JRRD.2012.09.0176",
language = "English (US)",
volume = "50",
pages = "997--1006",
journal = "Journal of rehabilitation R&D",
issn = "0007-506X",
publisher = "Department of Veterans Affairs",
number = "7",

}

TY - JOUR

T1 - OtoID

T2 - New extended frequency, portable audiometer for ototoxicity monitoring

AU - Dille, Marilyn F.

AU - Jacobs, Peter

AU - Gordon, Samuel Y.

AU - Helt, Wendy J.

AU - McMillan, Garnett P.

PY - 2013

Y1 - 2013

N2 - Portability of equipment is an increasingly important component in the practice of audiology. We report on a new device, the OtoID, that supports evidence-based ototoxicity testing protocols, provides capability for hearing testing on the hospital treatment unit, and can automate patient self-testing. The purpose of this article is to report on the validation and verification of the OtoID portable audiometer in 40 subjects both young and old, with and without hearing impairment. Subjects were evaluated by an audiologist using the manual hearing test program and then self-tested via an automated testing program. Testing was done in a sound booth and on a hospital treatment unit. Therefore, data were collected in four conditions (booth vs hospital unit and automated vs manual testing) and analyzed for testing bias, repeatability, and American Speech-Language-Hearing Association-significant ototoxicity false-positive rate. Repeatable hearing threshold results were obtained on all subjects who performed the test, regardless of hearing status or testing location.

AB - Portability of equipment is an increasingly important component in the practice of audiology. We report on a new device, the OtoID, that supports evidence-based ototoxicity testing protocols, provides capability for hearing testing on the hospital treatment unit, and can automate patient self-testing. The purpose of this article is to report on the validation and verification of the OtoID portable audiometer in 40 subjects both young and old, with and without hearing impairment. Subjects were evaluated by an audiologist using the manual hearing test program and then self-tested via an automated testing program. Testing was done in a sound booth and on a hospital treatment unit. Therefore, data were collected in four conditions (booth vs hospital unit and automated vs manual testing) and analyzed for testing bias, repeatability, and American Speech-Language-Hearing Association-significant ototoxicity false-positive rate. Repeatable hearing threshold results were obtained on all subjects who performed the test, regardless of hearing status or testing location.

KW - Audiometer

KW - Chemotherapy

KW - Cisplatin

KW - Hearing loss

KW - Monitoring

KW - OtoID

KW - Ototoxicity

KW - Rehabilitation

KW - Tinnitus

KW - Veterans

UR - http://www.scopus.com/inward/record.url?scp=84887300393&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84887300393&partnerID=8YFLogxK

U2 - 10.1682/JRRD.2012.09.0176

DO - 10.1682/JRRD.2012.09.0176

M3 - Article

C2 - 24301436

AN - SCOPUS:84887300393

VL - 50

SP - 997

EP - 1006

JO - Journal of rehabilitation R&D

JF - Journal of rehabilitation R&D

SN - 0007-506X

IS - 7

ER -