In vitro methods for studying the accuracy of velocity determination and spatial resolution of a color Doppler flow mapping system

Tadashi Tamura, Ajit Yoganathan, David Sahn

Research output: Contribution to journalArticle

31 Citations (Scopus)

Abstract

We tested a color Doppler flow mapping system (Aloka 880, 3.5 mHz) to image the steady flow of a suspension of cornstarch (0.1% to 1.75%) in water, flowing at known velocities through a series of Tygon tubing with internal diameters of 1.6 to 4.8 mm and latex rubber tubing (7.9 to 15.9 mm) in a water tank. Flow through the tubes was imaged at 2 to 12 cm and 2 to 18 cm distances from the transducer, by the use of the near and far focus settings of the instrument. First, the color flow diameters were compared to the known internal diameters of the tubing oriented at a 30- to 70- degree flow angle to the direction of interrogation to test axial resolution as well as laterel resolution. For the near transmit focus setting, the color diameters increasingly overestimated the actual flow diameters as the distance from the transducer increased progressively beyond 6 cm (i.e., overestimation of up to 25 mm or 500% for the 4.8 mm tubing and 250% for the 9.5 mm tubing at 16 cm from the transducer). For the far transmit focus setting and for distances within 10 cm, the color diameters did not differ significantly (i.e., more than 1 mm) from the actual diameter. The lowest velocity color encoded by the instrument was 3 cm/sec for a pulse repetition frequency of 4 kHz. At a flow velocity of about 50 cm/sec, 6 cm distance from the transducer and mid gain settings, flow through the 1.6 mm tube barely color coded, while flow thorugh 2.4, 3.2, and 4.8 mm tubes was easily imaged. These results have direct implications in clinical studies for imaging small septal defects, small orifices, or when measuring color flow Doppler diameters of vessels or valves for volume flow calculations. They also established a method for objectively evaluating the evolving capabilities of this new imaging technology.

Original languageEnglish (US)
Pages (from-to)152-158
Number of pages7
JournalAmerican Heart Journal
Volume114
Issue number1 PART 1
DOIs
StatePublished - 1987
Externally publishedYes

Fingerprint

Color
Transducers
Water
Rubber
In Vitro Techniques
Starch
Suspensions
Technology

ASJC Scopus subject areas

  • Cardiology and Cardiovascular Medicine

Cite this

In vitro methods for studying the accuracy of velocity determination and spatial resolution of a color Doppler flow mapping system. / Tamura, Tadashi; Yoganathan, Ajit; Sahn, David.

In: American Heart Journal, Vol. 114, No. 1 PART 1, 1987, p. 152-158.

Research output: Contribution to journalArticle

Tamura, Tadashi ; Yoganathan, Ajit ; Sahn, David. / In vitro methods for studying the accuracy of velocity determination and spatial resolution of a color Doppler flow mapping system. In: American Heart Journal. 1987 ; Vol. 114, No. 1 PART 1. pp. 152-158.
@article{59c4d7281c914e9b95a6fb258c7446c4,
title = "In vitro methods for studying the accuracy of velocity determination and spatial resolution of a color Doppler flow mapping system",
abstract = "We tested a color Doppler flow mapping system (Aloka 880, 3.5 mHz) to image the steady flow of a suspension of cornstarch (0.1{\%} to 1.75{\%}) in water, flowing at known velocities through a series of Tygon tubing with internal diameters of 1.6 to 4.8 mm and latex rubber tubing (7.9 to 15.9 mm) in a water tank. Flow through the tubes was imaged at 2 to 12 cm and 2 to 18 cm distances from the transducer, by the use of the near and far focus settings of the instrument. First, the color flow diameters were compared to the known internal diameters of the tubing oriented at a 30- to 70- degree flow angle to the direction of interrogation to test axial resolution as well as laterel resolution. For the near transmit focus setting, the color diameters increasingly overestimated the actual flow diameters as the distance from the transducer increased progressively beyond 6 cm (i.e., overestimation of up to 25 mm or 500{\%} for the 4.8 mm tubing and 250{\%} for the 9.5 mm tubing at 16 cm from the transducer). For the far transmit focus setting and for distances within 10 cm, the color diameters did not differ significantly (i.e., more than 1 mm) from the actual diameter. The lowest velocity color encoded by the instrument was 3 cm/sec for a pulse repetition frequency of 4 kHz. At a flow velocity of about 50 cm/sec, 6 cm distance from the transducer and mid gain settings, flow through the 1.6 mm tube barely color coded, while flow thorugh 2.4, 3.2, and 4.8 mm tubes was easily imaged. These results have direct implications in clinical studies for imaging small septal defects, small orifices, or when measuring color flow Doppler diameters of vessels or valves for volume flow calculations. They also established a method for objectively evaluating the evolving capabilities of this new imaging technology.",
author = "Tadashi Tamura and Ajit Yoganathan and David Sahn",
year = "1987",
doi = "10.1016/0002-8703(87)90319-X",
language = "English (US)",
volume = "114",
pages = "152--158",
journal = "American Heart Journal",
issn = "0002-8703",
publisher = "Mosby Inc.",
number = "1 PART 1",

}

TY - JOUR

T1 - In vitro methods for studying the accuracy of velocity determination and spatial resolution of a color Doppler flow mapping system

AU - Tamura, Tadashi

AU - Yoganathan, Ajit

AU - Sahn, David

PY - 1987

Y1 - 1987

N2 - We tested a color Doppler flow mapping system (Aloka 880, 3.5 mHz) to image the steady flow of a suspension of cornstarch (0.1% to 1.75%) in water, flowing at known velocities through a series of Tygon tubing with internal diameters of 1.6 to 4.8 mm and latex rubber tubing (7.9 to 15.9 mm) in a water tank. Flow through the tubes was imaged at 2 to 12 cm and 2 to 18 cm distances from the transducer, by the use of the near and far focus settings of the instrument. First, the color flow diameters were compared to the known internal diameters of the tubing oriented at a 30- to 70- degree flow angle to the direction of interrogation to test axial resolution as well as laterel resolution. For the near transmit focus setting, the color diameters increasingly overestimated the actual flow diameters as the distance from the transducer increased progressively beyond 6 cm (i.e., overestimation of up to 25 mm or 500% for the 4.8 mm tubing and 250% for the 9.5 mm tubing at 16 cm from the transducer). For the far transmit focus setting and for distances within 10 cm, the color diameters did not differ significantly (i.e., more than 1 mm) from the actual diameter. The lowest velocity color encoded by the instrument was 3 cm/sec for a pulse repetition frequency of 4 kHz. At a flow velocity of about 50 cm/sec, 6 cm distance from the transducer and mid gain settings, flow through the 1.6 mm tube barely color coded, while flow thorugh 2.4, 3.2, and 4.8 mm tubes was easily imaged. These results have direct implications in clinical studies for imaging small septal defects, small orifices, or when measuring color flow Doppler diameters of vessels or valves for volume flow calculations. They also established a method for objectively evaluating the evolving capabilities of this new imaging technology.

AB - We tested a color Doppler flow mapping system (Aloka 880, 3.5 mHz) to image the steady flow of a suspension of cornstarch (0.1% to 1.75%) in water, flowing at known velocities through a series of Tygon tubing with internal diameters of 1.6 to 4.8 mm and latex rubber tubing (7.9 to 15.9 mm) in a water tank. Flow through the tubes was imaged at 2 to 12 cm and 2 to 18 cm distances from the transducer, by the use of the near and far focus settings of the instrument. First, the color flow diameters were compared to the known internal diameters of the tubing oriented at a 30- to 70- degree flow angle to the direction of interrogation to test axial resolution as well as laterel resolution. For the near transmit focus setting, the color diameters increasingly overestimated the actual flow diameters as the distance from the transducer increased progressively beyond 6 cm (i.e., overestimation of up to 25 mm or 500% for the 4.8 mm tubing and 250% for the 9.5 mm tubing at 16 cm from the transducer). For the far transmit focus setting and for distances within 10 cm, the color diameters did not differ significantly (i.e., more than 1 mm) from the actual diameter. The lowest velocity color encoded by the instrument was 3 cm/sec for a pulse repetition frequency of 4 kHz. At a flow velocity of about 50 cm/sec, 6 cm distance from the transducer and mid gain settings, flow through the 1.6 mm tube barely color coded, while flow thorugh 2.4, 3.2, and 4.8 mm tubes was easily imaged. These results have direct implications in clinical studies for imaging small septal defects, small orifices, or when measuring color flow Doppler diameters of vessels or valves for volume flow calculations. They also established a method for objectively evaluating the evolving capabilities of this new imaging technology.

UR - http://www.scopus.com/inward/record.url?scp=0023236948&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0023236948&partnerID=8YFLogxK

U2 - 10.1016/0002-8703(87)90319-X

DO - 10.1016/0002-8703(87)90319-X

M3 - Article

C2 - 2955684

AN - SCOPUS:0023236948

VL - 114

SP - 152

EP - 158

JO - American Heart Journal

JF - American Heart Journal

SN - 0002-8703

IS - 1 PART 1

ER -