Foreign bodies in small arteries after use of an infusion microcatheter

Stanley L. Barnwell, Anthony N. D'Agostino, Steven L. Shapiro, Gary Nesbit, Jordi X. Kellogg

Research output: Contribution to journalArticle

47 Citations (Scopus)

Abstract

Over a 31-month period, we performed four neurointerventional procedures after which unexpected foreign bodies were noted in multiple arteries. All four procedures had in common the use of Fastracker-18 infusion microcatheters. Histologically, the intravascular debris looked strikingly similar to the hydrophilic coating on the catheter. An in vitro test mimicking clinical use of the microcatheter revealed that the hydrophilic coating can separate from the catheter. Until the coating is refined to make it more resistant to stripping, it may be advisable to reduce the amount of back-and-forth movement of these microcatheters if they have been positioned through guide catheters with small inner diameters and angled tips.

Original languageEnglish (US)
Pages (from-to)1886-1889
Number of pages4
JournalAmerican Journal of Neuroradiology
Volume18
Issue number10
StatePublished - Nov 1997

Fingerprint

Foreign Bodies
Catheters
Arteries

Keywords

  • Catheters and catheterization, complications
  • Foreign bodies
  • Latrogenic disease or disorder

ASJC Scopus subject areas

  • Clinical Neurology
  • Radiology Nuclear Medicine and imaging
  • Radiological and Ultrasound Technology

Cite this

Barnwell, S. L., D'Agostino, A. N., Shapiro, S. L., Nesbit, G., & Kellogg, J. X. (1997). Foreign bodies in small arteries after use of an infusion microcatheter. American Journal of Neuroradiology, 18(10), 1886-1889.

Foreign bodies in small arteries after use of an infusion microcatheter. / Barnwell, Stanley L.; D'Agostino, Anthony N.; Shapiro, Steven L.; Nesbit, Gary; Kellogg, Jordi X.

In: American Journal of Neuroradiology, Vol. 18, No. 10, 11.1997, p. 1886-1889.

Research output: Contribution to journalArticle

Barnwell, SL, D'Agostino, AN, Shapiro, SL, Nesbit, G & Kellogg, JX 1997, 'Foreign bodies in small arteries after use of an infusion microcatheter', American Journal of Neuroradiology, vol. 18, no. 10, pp. 1886-1889.
Barnwell SL, D'Agostino AN, Shapiro SL, Nesbit G, Kellogg JX. Foreign bodies in small arteries after use of an infusion microcatheter. American Journal of Neuroradiology. 1997 Nov;18(10):1886-1889.
Barnwell, Stanley L. ; D'Agostino, Anthony N. ; Shapiro, Steven L. ; Nesbit, Gary ; Kellogg, Jordi X. / Foreign bodies in small arteries after use of an infusion microcatheter. In: American Journal of Neuroradiology. 1997 ; Vol. 18, No. 10. pp. 1886-1889.
@article{b359535a7805426a9ff6121d6435079b,
title = "Foreign bodies in small arteries after use of an infusion microcatheter",
abstract = "Over a 31-month period, we performed four neurointerventional procedures after which unexpected foreign bodies were noted in multiple arteries. All four procedures had in common the use of Fastracker-18 infusion microcatheters. Histologically, the intravascular debris looked strikingly similar to the hydrophilic coating on the catheter. An in vitro test mimicking clinical use of the microcatheter revealed that the hydrophilic coating can separate from the catheter. Until the coating is refined to make it more resistant to stripping, it may be advisable to reduce the amount of back-and-forth movement of these microcatheters if they have been positioned through guide catheters with small inner diameters and angled tips.",
keywords = "Catheters and catheterization, complications, Foreign bodies, Latrogenic disease or disorder",
author = "Barnwell, {Stanley L.} and D'Agostino, {Anthony N.} and Shapiro, {Steven L.} and Gary Nesbit and Kellogg, {Jordi X.}",
year = "1997",
month = "11",
language = "English (US)",
volume = "18",
pages = "1886--1889",
journal = "American Journal of Neuroradiology",
issn = "0195-6108",
publisher = "American Society of Neuroradiology",
number = "10",

}

TY - JOUR

T1 - Foreign bodies in small arteries after use of an infusion microcatheter

AU - Barnwell, Stanley L.

AU - D'Agostino, Anthony N.

AU - Shapiro, Steven L.

AU - Nesbit, Gary

AU - Kellogg, Jordi X.

PY - 1997/11

Y1 - 1997/11

N2 - Over a 31-month period, we performed four neurointerventional procedures after which unexpected foreign bodies were noted in multiple arteries. All four procedures had in common the use of Fastracker-18 infusion microcatheters. Histologically, the intravascular debris looked strikingly similar to the hydrophilic coating on the catheter. An in vitro test mimicking clinical use of the microcatheter revealed that the hydrophilic coating can separate from the catheter. Until the coating is refined to make it more resistant to stripping, it may be advisable to reduce the amount of back-and-forth movement of these microcatheters if they have been positioned through guide catheters with small inner diameters and angled tips.

AB - Over a 31-month period, we performed four neurointerventional procedures after which unexpected foreign bodies were noted in multiple arteries. All four procedures had in common the use of Fastracker-18 infusion microcatheters. Histologically, the intravascular debris looked strikingly similar to the hydrophilic coating on the catheter. An in vitro test mimicking clinical use of the microcatheter revealed that the hydrophilic coating can separate from the catheter. Until the coating is refined to make it more resistant to stripping, it may be advisable to reduce the amount of back-and-forth movement of these microcatheters if they have been positioned through guide catheters with small inner diameters and angled tips.

KW - Catheters and catheterization, complications

KW - Foreign bodies

KW - Latrogenic disease or disorder

UR - http://www.scopus.com/inward/record.url?scp=0031546003&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0031546003&partnerID=8YFLogxK

M3 - Article

VL - 18

SP - 1886

EP - 1889