T1 - Effect of pupil dilation on biometric measurements and intraocular lens power calculations in schoolchildren
AU - Wang, Xiaogang
AU - Dong, Jing
AU - Tang, Maolong
AU - Wang, Xiaoliang
AU - Wang, Haitao
AU - Zhang, Suhua
N1 - Publisher Copyright:
© 2018 Wang et al. This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
PY - 2018/9
Y1 - 2018/9
N2 - Purpose To investigate the effect of pupil dilation on ocular biometric parameters and intraocular lens (IOL) power calculation in schoolchildren using the Lenstar LS 900. Methods One hundred forty eyes of 140 healthy schoolchildren were included in the analysis. Axial length (AL), central corneal thickness (CCT), aqueous depth (AD), anterior chamber depth (ACD), lens thickness (LT), flat keratometry (K), steep K, astigmatism, white-to-white (WTW), and iris/pupil barycenter distance were measured, before and after pupil dilation. Anterior segment length (ASL) was defined as the sum of ACD and LT, and lens position (LP) was defined as ACD plus half of the LT. The relative lens position (RLP) was defined as LP divided by AL. IOL power was calculated using the eight formulas (Hill-RBF, Barrett, Haigis, Hoffer Q, Holladay, Olsen, SRK II, and SRK/T) integrated in the Lenstar LS 900. Parameters before and after pupil dilation were compared. Results AL, AD, ACD, LT, ASL, LP, RLP, flat K, iris barycenter distance, pupil barycenter distance, and PD differed significantly after pupil dilation (P < 0.001 in all cases), as compared to before dilation. The Olsen formula demonstrated significant differences in the magnitude of astigmatism (P = 0.010) and IOL power (P = 0.003) after pupil dilation. Using the different formulas, 23.6–40.7% of participants had IOL power changes of more than 0.50 diopters, while 0.7–1.4% had IOL changes of more than 1.0 diopter after pupil dilation. Conclusions Dilated and undilated pupil size affected the Lenstar LS 900 measurement of some ocular biometric parameters, and pupil dilation led to IOL power changes exceeding 0.50 diopters with a high percentage (from 23.6% to 40.7%) in schoolchildren, which should be noticed in clinical practice.
AB - Purpose To investigate the effect of pupil dilation on ocular biometric parameters and intraocular lens (IOL) power calculation in schoolchildren using the Lenstar LS 900. Methods One hundred forty eyes of 140 healthy schoolchildren were included in the analysis. Axial length (AL), central corneal thickness (CCT), aqueous depth (AD), anterior chamber depth (ACD), lens thickness (LT), flat keratometry (K), steep K, astigmatism, white-to-white (WTW), and iris/pupil barycenter distance were measured, before and after pupil dilation. Anterior segment length (ASL) was defined as the sum of ACD and LT, and lens position (LP) was defined as ACD plus half of the LT. The relative lens position (RLP) was defined as LP divided by AL. IOL power was calculated using the eight formulas (Hill-RBF, Barrett, Haigis, Hoffer Q, Holladay, Olsen, SRK II, and SRK/T) integrated in the Lenstar LS 900. Parameters before and after pupil dilation were compared. Results AL, AD, ACD, LT, ASL, LP, RLP, flat K, iris barycenter distance, pupil barycenter distance, and PD differed significantly after pupil dilation (P < 0.001 in all cases), as compared to before dilation. The Olsen formula demonstrated significant differences in the magnitude of astigmatism (P = 0.010) and IOL power (P = 0.003) after pupil dilation. Using the different formulas, 23.6–40.7% of participants had IOL power changes of more than 0.50 diopters, while 0.7–1.4% had IOL changes of more than 1.0 diopter after pupil dilation. Conclusions Dilated and undilated pupil size affected the Lenstar LS 900 measurement of some ocular biometric parameters, and pupil dilation led to IOL power changes exceeding 0.50 diopters with a high percentage (from 23.6% to 40.7%) in schoolchildren, which should be noticed in clinical practice.
UR - http://www.scopus.com/inward/record.url?scp=85053271785&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85053271785&partnerID=8YFLogxK
U2 - 10.1371/journal.pone.0203677
DO - 10.1371/journal.pone.0203677
M3 - Article
C2 - 30212545
AN - SCOPUS:85053271785
SN - 1932-6203
VL - 13
JO - PloS one
JF - PloS one
IS - 9
M1 - e0203677
ER -