A system to measure thermal nociception

James Galbraith, Bruce J. Mrosko, Robert R. Myers

Research output: Contribution to journalArticle

33 Citations (Scopus)

Abstract

Hyperalgesia is commonly demonstrated by measuring the difference in latency to thermal nociception between a control and injured condition. Historically these tests have been conducted using either a hot plate or radiant heat source to stimulate thermal receptors in the foot. The design of a simple and inexpensive circuit to measure thermal nociception based on a method originally introduced by Hargreaves et al. (1988) is described. The design uses a movable, high-intensity light source to heat either the right or left footpad of the test animal. The advantage of this method is that it permits each animal to serve as its own control since the experimental and control footpads can be tested independently. The design possesses greater sensitivity than mechanical hyperalgesia test procedures and provides an automated endpoint that electronically records the latency to removal of the foot from the heat source. Additional features of this design which reduce variability include a constant temperature floor to reduce the effect of altered footpad temperature that may be caused by autonomic dysfunction in the experimental limb and a method of reproducibly locating the heat source prior to testing. The response to a unilateral lesion is demonstrated with this device.

Original languageEnglish (US)
Pages (from-to)63-68
Number of pages6
JournalJournal of Neuroscience Methods
Volume49
Issue number1-2
DOIs
StatePublished - 1993
Externally publishedYes

Fingerprint

Nociception
Hot Temperature
Hyperalgesia
Foot
Temperature
Extremities
Light
Equipment and Supplies

Keywords

  • Behavioral hyperalgesia
  • Painful neuropathy
  • Thermal nociception

ASJC Scopus subject areas

  • Neuroscience(all)

Cite this

A system to measure thermal nociception. / Galbraith, James; Mrosko, Bruce J.; Myers, Robert R.

In: Journal of Neuroscience Methods, Vol. 49, No. 1-2, 1993, p. 63-68.

Research output: Contribution to journalArticle

Galbraith, James ; Mrosko, Bruce J. ; Myers, Robert R. / A system to measure thermal nociception. In: Journal of Neuroscience Methods. 1993 ; Vol. 49, No. 1-2. pp. 63-68.
@article{a12cee6e2fbd414fb161c42c124a382d,
title = "A system to measure thermal nociception",
abstract = "Hyperalgesia is commonly demonstrated by measuring the difference in latency to thermal nociception between a control and injured condition. Historically these tests have been conducted using either a hot plate or radiant heat source to stimulate thermal receptors in the foot. The design of a simple and inexpensive circuit to measure thermal nociception based on a method originally introduced by Hargreaves et al. (1988) is described. The design uses a movable, high-intensity light source to heat either the right or left footpad of the test animal. The advantage of this method is that it permits each animal to serve as its own control since the experimental and control footpads can be tested independently. The design possesses greater sensitivity than mechanical hyperalgesia test procedures and provides an automated endpoint that electronically records the latency to removal of the foot from the heat source. Additional features of this design which reduce variability include a constant temperature floor to reduce the effect of altered footpad temperature that may be caused by autonomic dysfunction in the experimental limb and a method of reproducibly locating the heat source prior to testing. The response to a unilateral lesion is demonstrated with this device.",
keywords = "Behavioral hyperalgesia, Painful neuropathy, Thermal nociception",
author = "James Galbraith and Mrosko, {Bruce J.} and Myers, {Robert R.}",
year = "1993",
doi = "10.1016/0165-0270(93)90109-5",
language = "English (US)",
volume = "49",
pages = "63--68",
journal = "Journal of Neuroscience Methods",
issn = "0165-0270",
publisher = "Elsevier",
number = "1-2",

}

TY - JOUR

T1 - A system to measure thermal nociception

AU - Galbraith, James

AU - Mrosko, Bruce J.

AU - Myers, Robert R.

PY - 1993

Y1 - 1993

N2 - Hyperalgesia is commonly demonstrated by measuring the difference in latency to thermal nociception between a control and injured condition. Historically these tests have been conducted using either a hot plate or radiant heat source to stimulate thermal receptors in the foot. The design of a simple and inexpensive circuit to measure thermal nociception based on a method originally introduced by Hargreaves et al. (1988) is described. The design uses a movable, high-intensity light source to heat either the right or left footpad of the test animal. The advantage of this method is that it permits each animal to serve as its own control since the experimental and control footpads can be tested independently. The design possesses greater sensitivity than mechanical hyperalgesia test procedures and provides an automated endpoint that electronically records the latency to removal of the foot from the heat source. Additional features of this design which reduce variability include a constant temperature floor to reduce the effect of altered footpad temperature that may be caused by autonomic dysfunction in the experimental limb and a method of reproducibly locating the heat source prior to testing. The response to a unilateral lesion is demonstrated with this device.

AB - Hyperalgesia is commonly demonstrated by measuring the difference in latency to thermal nociception between a control and injured condition. Historically these tests have been conducted using either a hot plate or radiant heat source to stimulate thermal receptors in the foot. The design of a simple and inexpensive circuit to measure thermal nociception based on a method originally introduced by Hargreaves et al. (1988) is described. The design uses a movable, high-intensity light source to heat either the right or left footpad of the test animal. The advantage of this method is that it permits each animal to serve as its own control since the experimental and control footpads can be tested independently. The design possesses greater sensitivity than mechanical hyperalgesia test procedures and provides an automated endpoint that electronically records the latency to removal of the foot from the heat source. Additional features of this design which reduce variability include a constant temperature floor to reduce the effect of altered footpad temperature that may be caused by autonomic dysfunction in the experimental limb and a method of reproducibly locating the heat source prior to testing. The response to a unilateral lesion is demonstrated with this device.

KW - Behavioral hyperalgesia

KW - Painful neuropathy

KW - Thermal nociception

UR - http://www.scopus.com/inward/record.url?scp=0027318022&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0027318022&partnerID=8YFLogxK

U2 - 10.1016/0165-0270(93)90109-5

DO - 10.1016/0165-0270(93)90109-5

M3 - Article

C2 - 8271832

AN - SCOPUS:0027318022

VL - 49

SP - 63

EP - 68

JO - Journal of Neuroscience Methods

JF - Journal of Neuroscience Methods

SN - 0165-0270

IS - 1-2

ER -