The volume effect of detectors in the dosimetry of small fields used in IMRT

Wolfram Laub, Tony Wong

Research output: Contribution to journalArticle

210 Citations (Scopus)

Abstract

In this study we investigate the effect of detector size in the dosimetry of small fields and steep dose gradients with a particular emphasis on IMRT measurements. Comparisons of calculated and measured cross-profiles and absolute dose values of IMRT treatment plans are presented. As a consequence of the finite size of the detector that was used for the commissioning of the IMRT tool, local discrepancies of more than 10% are found between calculated cross-profiles of intensity modulated beams and intensity modulated profiles measured with film. Absolute dose measurements of intensity modulated fields with a 0.6 cm3 Farmer chamber show significant differences of more than 6% between calculated and measured dose values at the isocenter of an IMRT treatment plan, Differences of not more than 2% are found in the same experiment for dose values measured with a 0.015 cm3 pinpoint ion chamber. A method to correct for the spatial response of finite-sized detectors and to obtain the "real" penumbra width of cross-profiles from measurements is introduced. Output factor measurements are performed with different detectors and are presented as a function of detector size for a 1 × 1 cm2 field. Because of its high spatial resolution and water equivalence, a diamond detector is found to be suitable as an alternative to other detectors used for small field dosimetry as there are photographic and photochromic film, TLDs, or water-equivalent scintillation detectors.

Original languageEnglish (US)
Pages (from-to)341-347
Number of pages7
JournalMedical Physics
Volume30
Issue number3
DOIs
StatePublished - Mar 1 2003
Externally publishedYes

Fingerprint

Diamond
Water
Ions
Therapeutics
Farmers

Keywords

  • IMRT
  • Quality assurance
  • Small field dosimetry
  • Volume effect

ASJC Scopus subject areas

  • Biophysics
  • Radiology Nuclear Medicine and imaging

Cite this

The volume effect of detectors in the dosimetry of small fields used in IMRT. / Laub, Wolfram; Wong, Tony.

In: Medical Physics, Vol. 30, No. 3, 01.03.2003, p. 341-347.

Research output: Contribution to journalArticle

Laub, Wolfram ; Wong, Tony. / The volume effect of detectors in the dosimetry of small fields used in IMRT. In: Medical Physics. 2003 ; Vol. 30, No. 3. pp. 341-347.
@article{6a2ff50bd50f4a5ca1388fd97ec04bf8,
title = "The volume effect of detectors in the dosimetry of small fields used in IMRT",
abstract = "In this study we investigate the effect of detector size in the dosimetry of small fields and steep dose gradients with a particular emphasis on IMRT measurements. Comparisons of calculated and measured cross-profiles and absolute dose values of IMRT treatment plans are presented. As a consequence of the finite size of the detector that was used for the commissioning of the IMRT tool, local discrepancies of more than 10{\%} are found between calculated cross-profiles of intensity modulated beams and intensity modulated profiles measured with film. Absolute dose measurements of intensity modulated fields with a 0.6 cm3 Farmer chamber show significant differences of more than 6{\%} between calculated and measured dose values at the isocenter of an IMRT treatment plan, Differences of not more than 2{\%} are found in the same experiment for dose values measured with a 0.015 cm3 pinpoint ion chamber. A method to correct for the spatial response of finite-sized detectors and to obtain the {"}real{"} penumbra width of cross-profiles from measurements is introduced. Output factor measurements are performed with different detectors and are presented as a function of detector size for a 1 × 1 cm2 field. Because of its high spatial resolution and water equivalence, a diamond detector is found to be suitable as an alternative to other detectors used for small field dosimetry as there are photographic and photochromic film, TLDs, or water-equivalent scintillation detectors.",
keywords = "IMRT, Quality assurance, Small field dosimetry, Volume effect",
author = "Wolfram Laub and Tony Wong",
year = "2003",
month = "3",
day = "1",
doi = "10.1118/1.1544678",
language = "English (US)",
volume = "30",
pages = "341--347",
journal = "Medical Physics",
issn = "0094-2405",
publisher = "AAPM - American Association of Physicists in Medicine",
number = "3",

}

TY - JOUR

T1 - The volume effect of detectors in the dosimetry of small fields used in IMRT

AU - Laub, Wolfram

AU - Wong, Tony

PY - 2003/3/1

Y1 - 2003/3/1

N2 - In this study we investigate the effect of detector size in the dosimetry of small fields and steep dose gradients with a particular emphasis on IMRT measurements. Comparisons of calculated and measured cross-profiles and absolute dose values of IMRT treatment plans are presented. As a consequence of the finite size of the detector that was used for the commissioning of the IMRT tool, local discrepancies of more than 10% are found between calculated cross-profiles of intensity modulated beams and intensity modulated profiles measured with film. Absolute dose measurements of intensity modulated fields with a 0.6 cm3 Farmer chamber show significant differences of more than 6% between calculated and measured dose values at the isocenter of an IMRT treatment plan, Differences of not more than 2% are found in the same experiment for dose values measured with a 0.015 cm3 pinpoint ion chamber. A method to correct for the spatial response of finite-sized detectors and to obtain the "real" penumbra width of cross-profiles from measurements is introduced. Output factor measurements are performed with different detectors and are presented as a function of detector size for a 1 × 1 cm2 field. Because of its high spatial resolution and water equivalence, a diamond detector is found to be suitable as an alternative to other detectors used for small field dosimetry as there are photographic and photochromic film, TLDs, or water-equivalent scintillation detectors.

AB - In this study we investigate the effect of detector size in the dosimetry of small fields and steep dose gradients with a particular emphasis on IMRT measurements. Comparisons of calculated and measured cross-profiles and absolute dose values of IMRT treatment plans are presented. As a consequence of the finite size of the detector that was used for the commissioning of the IMRT tool, local discrepancies of more than 10% are found between calculated cross-profiles of intensity modulated beams and intensity modulated profiles measured with film. Absolute dose measurements of intensity modulated fields with a 0.6 cm3 Farmer chamber show significant differences of more than 6% between calculated and measured dose values at the isocenter of an IMRT treatment plan, Differences of not more than 2% are found in the same experiment for dose values measured with a 0.015 cm3 pinpoint ion chamber. A method to correct for the spatial response of finite-sized detectors and to obtain the "real" penumbra width of cross-profiles from measurements is introduced. Output factor measurements are performed with different detectors and are presented as a function of detector size for a 1 × 1 cm2 field. Because of its high spatial resolution and water equivalence, a diamond detector is found to be suitable as an alternative to other detectors used for small field dosimetry as there are photographic and photochromic film, TLDs, or water-equivalent scintillation detectors.

KW - IMRT

KW - Quality assurance

KW - Small field dosimetry

KW - Volume effect

UR - http://www.scopus.com/inward/record.url?scp=0037339773&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0037339773&partnerID=8YFLogxK

U2 - 10.1118/1.1544678

DO - 10.1118/1.1544678

M3 - Article

C2 - 12674234

AN - SCOPUS:0037339773

VL - 30

SP - 341

EP - 347

JO - Medical Physics

JF - Medical Physics

SN - 0094-2405

IS - 3

ER -