Reliability and accuracy of the thoracic impedance signal for measuring cardiopulmonary resuscitation quality metrics

Erik Alonso, Jesús Ruiz, Elisabete Aramendi, Digna González-Otero, Sofía Ruiz de Gauna, Unai Ayala, James K. Russell, Mohamud Ramzan Daya

Research output: Contribution to journalArticle

14 Citations (Scopus)

Abstract

Aim: To determine the accuracy and reliability of the thoracic impedance (TI) signal to assess cardiopulmonary resuscitation (CPR) quality metrics. Methods: A dataset of 63 out-of-hospital cardiac arrest episodes containing the compression depth (CD), capnography and TI signals was used. We developed a chest compression (CC) and ventilation detector based on the TI signal. TI shows fluctuations due to CCs and ventilations. A decision algorithm classified the local maxima as CCs or ventilations. Seven CPR quality metrics were computed: mean CC-rate, fraction of minutes with inadequate CC-rate, chest compression fraction, mean ventilation rate, fraction of minutes with hyperventilation, instantaneous CC-rate and instantaneous ventilation rate. The CD and capnography signals were accepted as the gold standard for CC and ventilation detection respectively. The accuracy of the detector was evaluated in terms of sensitivity and positive predictive value (PPV). Distributions for each metric computed from the TI and from the gold standard were calculated and tested for normality using one sample Kolmogorov-Smirnov test. For normal and not normal distributions, two sample t-test and Mann-Whitney U test respectively were applied to test for equal means and medians respectively. Bland-Altman plots were represented for each metric to analyze the level of agreement between values obtained from the TI and gold standard. Results: The CC/ventilation detector had a median sensitivity/PPV of 97.2%/97.7% for CCs and 92.2%/81.0% for ventilations respectively. Distributions for all the metrics showed equal means or medians, and agreements >95% between metrics and gold standard was achieved for most of the episodes in the test set, except for the instantaneous ventilation rate. Conclusion: With our data, the TI can be reliably used to measure all the CPR quality metrics proposed in this study, except for the instantaneous ventilation rate.

Original languageEnglish (US)
Pages (from-to)28-34
Number of pages7
JournalResuscitation
Volume88
DOIs
StatePublished - Mar 1 2015

Fingerprint

Cardiopulmonary Resuscitation
Electric Impedance
Thorax
Ventilation
Capnography
Nonparametric Statistics
Out-of-Hospital Cardiac Arrest
Hyperventilation
Normal Distribution

Keywords

  • Automated external defibrillator
  • Cardiopulmonary resuscitation quality
  • Chest compression
  • Thoracic impedance
  • Ventilations

ASJC Scopus subject areas

  • Cardiology and Cardiovascular Medicine
  • Emergency
  • Emergency Medicine

Cite this

Reliability and accuracy of the thoracic impedance signal for measuring cardiopulmonary resuscitation quality metrics. / Alonso, Erik; Ruiz, Jesús; Aramendi, Elisabete; González-Otero, Digna; Ruiz de Gauna, Sofía; Ayala, Unai; Russell, James K.; Daya, Mohamud Ramzan.

In: Resuscitation, Vol. 88, 01.03.2015, p. 28-34.

Research output: Contribution to journalArticle

Alonso, Erik ; Ruiz, Jesús ; Aramendi, Elisabete ; González-Otero, Digna ; Ruiz de Gauna, Sofía ; Ayala, Unai ; Russell, James K. ; Daya, Mohamud Ramzan. / Reliability and accuracy of the thoracic impedance signal for measuring cardiopulmonary resuscitation quality metrics. In: Resuscitation. 2015 ; Vol. 88. pp. 28-34.
@article{25b27d25d20d47958a2d38b3a3b3dbfe,
title = "Reliability and accuracy of the thoracic impedance signal for measuring cardiopulmonary resuscitation quality metrics",
abstract = "Aim: To determine the accuracy and reliability of the thoracic impedance (TI) signal to assess cardiopulmonary resuscitation (CPR) quality metrics. Methods: A dataset of 63 out-of-hospital cardiac arrest episodes containing the compression depth (CD), capnography and TI signals was used. We developed a chest compression (CC) and ventilation detector based on the TI signal. TI shows fluctuations due to CCs and ventilations. A decision algorithm classified the local maxima as CCs or ventilations. Seven CPR quality metrics were computed: mean CC-rate, fraction of minutes with inadequate CC-rate, chest compression fraction, mean ventilation rate, fraction of minutes with hyperventilation, instantaneous CC-rate and instantaneous ventilation rate. The CD and capnography signals were accepted as the gold standard for CC and ventilation detection respectively. The accuracy of the detector was evaluated in terms of sensitivity and positive predictive value (PPV). Distributions for each metric computed from the TI and from the gold standard were calculated and tested for normality using one sample Kolmogorov-Smirnov test. For normal and not normal distributions, two sample t-test and Mann-Whitney U test respectively were applied to test for equal means and medians respectively. Bland-Altman plots were represented for each metric to analyze the level of agreement between values obtained from the TI and gold standard. Results: The CC/ventilation detector had a median sensitivity/PPV of 97.2{\%}/97.7{\%} for CCs and 92.2{\%}/81.0{\%} for ventilations respectively. Distributions for all the metrics showed equal means or medians, and agreements >95{\%} between metrics and gold standard was achieved for most of the episodes in the test set, except for the instantaneous ventilation rate. Conclusion: With our data, the TI can be reliably used to measure all the CPR quality metrics proposed in this study, except for the instantaneous ventilation rate.",
keywords = "Automated external defibrillator, Cardiopulmonary resuscitation quality, Chest compression, Thoracic impedance, Ventilations",
author = "Erik Alonso and Jes{\'u}s Ruiz and Elisabete Aramendi and Digna Gonz{\'a}lez-Otero and {Ruiz de Gauna}, Sof{\'i}a and Unai Ayala and Russell, {James K.} and Daya, {Mohamud Ramzan}",
year = "2015",
month = "3",
day = "1",
doi = "10.1016/j.resuscitation.2014.11.027",
language = "English (US)",
volume = "88",
pages = "28--34",
journal = "Resuscitation",
issn = "0300-9572",
publisher = "Elsevier Ireland Ltd",

}

TY - JOUR

T1 - Reliability and accuracy of the thoracic impedance signal for measuring cardiopulmonary resuscitation quality metrics

AU - Alonso, Erik

AU - Ruiz, Jesús

AU - Aramendi, Elisabete

AU - González-Otero, Digna

AU - Ruiz de Gauna, Sofía

AU - Ayala, Unai

AU - Russell, James K.

AU - Daya, Mohamud Ramzan

PY - 2015/3/1

Y1 - 2015/3/1

N2 - Aim: To determine the accuracy and reliability of the thoracic impedance (TI) signal to assess cardiopulmonary resuscitation (CPR) quality metrics. Methods: A dataset of 63 out-of-hospital cardiac arrest episodes containing the compression depth (CD), capnography and TI signals was used. We developed a chest compression (CC) and ventilation detector based on the TI signal. TI shows fluctuations due to CCs and ventilations. A decision algorithm classified the local maxima as CCs or ventilations. Seven CPR quality metrics were computed: mean CC-rate, fraction of minutes with inadequate CC-rate, chest compression fraction, mean ventilation rate, fraction of minutes with hyperventilation, instantaneous CC-rate and instantaneous ventilation rate. The CD and capnography signals were accepted as the gold standard for CC and ventilation detection respectively. The accuracy of the detector was evaluated in terms of sensitivity and positive predictive value (PPV). Distributions for each metric computed from the TI and from the gold standard were calculated and tested for normality using one sample Kolmogorov-Smirnov test. For normal and not normal distributions, two sample t-test and Mann-Whitney U test respectively were applied to test for equal means and medians respectively. Bland-Altman plots were represented for each metric to analyze the level of agreement between values obtained from the TI and gold standard. Results: The CC/ventilation detector had a median sensitivity/PPV of 97.2%/97.7% for CCs and 92.2%/81.0% for ventilations respectively. Distributions for all the metrics showed equal means or medians, and agreements >95% between metrics and gold standard was achieved for most of the episodes in the test set, except for the instantaneous ventilation rate. Conclusion: With our data, the TI can be reliably used to measure all the CPR quality metrics proposed in this study, except for the instantaneous ventilation rate.

AB - Aim: To determine the accuracy and reliability of the thoracic impedance (TI) signal to assess cardiopulmonary resuscitation (CPR) quality metrics. Methods: A dataset of 63 out-of-hospital cardiac arrest episodes containing the compression depth (CD), capnography and TI signals was used. We developed a chest compression (CC) and ventilation detector based on the TI signal. TI shows fluctuations due to CCs and ventilations. A decision algorithm classified the local maxima as CCs or ventilations. Seven CPR quality metrics were computed: mean CC-rate, fraction of minutes with inadequate CC-rate, chest compression fraction, mean ventilation rate, fraction of minutes with hyperventilation, instantaneous CC-rate and instantaneous ventilation rate. The CD and capnography signals were accepted as the gold standard for CC and ventilation detection respectively. The accuracy of the detector was evaluated in terms of sensitivity and positive predictive value (PPV). Distributions for each metric computed from the TI and from the gold standard were calculated and tested for normality using one sample Kolmogorov-Smirnov test. For normal and not normal distributions, two sample t-test and Mann-Whitney U test respectively were applied to test for equal means and medians respectively. Bland-Altman plots were represented for each metric to analyze the level of agreement between values obtained from the TI and gold standard. Results: The CC/ventilation detector had a median sensitivity/PPV of 97.2%/97.7% for CCs and 92.2%/81.0% for ventilations respectively. Distributions for all the metrics showed equal means or medians, and agreements >95% between metrics and gold standard was achieved for most of the episodes in the test set, except for the instantaneous ventilation rate. Conclusion: With our data, the TI can be reliably used to measure all the CPR quality metrics proposed in this study, except for the instantaneous ventilation rate.

KW - Automated external defibrillator

KW - Cardiopulmonary resuscitation quality

KW - Chest compression

KW - Thoracic impedance

KW - Ventilations

UR - http://www.scopus.com/inward/record.url?scp=84922805049&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84922805049&partnerID=8YFLogxK

U2 - 10.1016/j.resuscitation.2014.11.027

DO - 10.1016/j.resuscitation.2014.11.027

M3 - Article

C2 - 25524362

AN - SCOPUS:84922805049

VL - 88

SP - 28

EP - 34

JO - Resuscitation

JF - Resuscitation

SN - 0300-9572

ER -