Phonocardiographic diagnosis of aortic ball variance.

J. C. Hylen, F. E. Kloster, R. H. Herr, P. Q. Hull, A. W. Ames, A. Starr, H. E. Griswold

Research output: Contribution to journalArticlepeer-review

46 Scopus citations

Abstract

Fatty infiltration causing changes in the silastic poppet of the Model 1000 series Starr-Edwards aortic valve prostheses (ball variance) has been detected with increasing frequency and can result in sudden death. Phonocardiograms were recorded on 12 patients with ball variance confirmed by operation and of 31 controls. Ten of the 12 patients with ball variance were distinguished from the controls by an aortic opening sound (AO) less than half as intense as the aortic closure sound (AC) at the second right intercostal space (AO/AC ratio less than 0.5). Both AO and AC were decreased in two patients with ball variance, with the loss of the characteristic high frequency and amplitude of these sounds. The only patient having a diminished AO/AC ratio (0.42) without ball variance at reoperation had a clot extending over the aortic valve struts. The phonocardiographic findings have been the most reliable objective evidence of ball variance in patients with Starr-Edwards aortic prosthesis of the Model 1000 series.