Noninvasive measurement of rat intraocular pressure with the Tono-Pen

C. G. Moore, S. T. Milne, John Morrison

Research output: Contribution to journalArticle

139 Citations (Scopus)

Abstract

Purpose. The purpose of this study was to evaluate the Tono-Pen 2 tonometer for measuring intraocular pressure (IOP) in the living rat eye. Methods. One eye from each of 20 adult, anesthetized brown Norway rats (group 1) was cannulated and simultaneously connected to a syringe and a pressure transducer with a chart recorder. We increased IOP from 15 to 45 mmHg in 5- mmHg increments and obtained 15 consecutive readings (ignoring instrument- generated averages) at each pressure increment with a Tono-Pen 2 tonometer. To test the tonopen's ability to measure unknown IOP, transducer pressures were varied randomly in 20 additional animals (group 2), and tonopen readings were obtained in masked fashion. Results. Plotting the mean tonopen readings for each animal against transducer IOP produced a regression formula of y = 4.54 + 0.79x (r = 0.98). Mean group 2 tonopen values plotted against transducer IOP yielded a regression formula of y = 4.75 + 0.78x (r = 0.94). A method comparison analysis showed that the tonopen significantly overestimates pressures at low IOP (≤ 15 mmHg), and it significantly underestimates pressures at high IOP (≥ 30 mmHg). Using two-way analysis of variance, it was determined that the group 2 data did not differ significantly from the group 1 data (P ≥ 0.76). Because of this consistency, we generated a correction factor with 95% prediction intervals for Tono-Pen readings. Conclusions. The Tono-Pen 2 can be used reliably to measure IOP in the normal rat eye.

Original languageEnglish (US)
Pages (from-to)363-369
Number of pages7
JournalInvestigative Ophthalmology and Visual Science
Volume34
Issue number2
StatePublished - 1993

Fingerprint

Intraocular Pressure
Reading
Pressure Transducers
Pressure
Transducers
Syringes
Analysis of Variance

Keywords

  • intraocular pressure
  • rat
  • Tono-Pen
  • tonometry

ASJC Scopus subject areas

  • Ophthalmology

Cite this

Noninvasive measurement of rat intraocular pressure with the Tono-Pen. / Moore, C. G.; Milne, S. T.; Morrison, John.

In: Investigative Ophthalmology and Visual Science, Vol. 34, No. 2, 1993, p. 363-369.

Research output: Contribution to journalArticle

@article{18abec1447644efd85fa295fe8737f09,
title = "Noninvasive measurement of rat intraocular pressure with the Tono-Pen",
abstract = "Purpose. The purpose of this study was to evaluate the Tono-Pen 2 tonometer for measuring intraocular pressure (IOP) in the living rat eye. Methods. One eye from each of 20 adult, anesthetized brown Norway rats (group 1) was cannulated and simultaneously connected to a syringe and a pressure transducer with a chart recorder. We increased IOP from 15 to 45 mmHg in 5- mmHg increments and obtained 15 consecutive readings (ignoring instrument- generated averages) at each pressure increment with a Tono-Pen 2 tonometer. To test the tonopen's ability to measure unknown IOP, transducer pressures were varied randomly in 20 additional animals (group 2), and tonopen readings were obtained in masked fashion. Results. Plotting the mean tonopen readings for each animal against transducer IOP produced a regression formula of y = 4.54 + 0.79x (r = 0.98). Mean group 2 tonopen values plotted against transducer IOP yielded a regression formula of y = 4.75 + 0.78x (r = 0.94). A method comparison analysis showed that the tonopen significantly overestimates pressures at low IOP (≤ 15 mmHg), and it significantly underestimates pressures at high IOP (≥ 30 mmHg). Using two-way analysis of variance, it was determined that the group 2 data did not differ significantly from the group 1 data (P ≥ 0.76). Because of this consistency, we generated a correction factor with 95{\%} prediction intervals for Tono-Pen readings. Conclusions. The Tono-Pen 2 can be used reliably to measure IOP in the normal rat eye.",
keywords = "intraocular pressure, rat, Tono-Pen, tonometry",
author = "Moore, {C. G.} and Milne, {S. T.} and John Morrison",
year = "1993",
language = "English (US)",
volume = "34",
pages = "363--369",
journal = "Investigative Ophthalmology and Visual Science",
issn = "0146-0404",
publisher = "Association for Research in Vision and Ophthalmology Inc.",
number = "2",

}

TY - JOUR

T1 - Noninvasive measurement of rat intraocular pressure with the Tono-Pen

AU - Moore, C. G.

AU - Milne, S. T.

AU - Morrison, John

PY - 1993

Y1 - 1993

N2 - Purpose. The purpose of this study was to evaluate the Tono-Pen 2 tonometer for measuring intraocular pressure (IOP) in the living rat eye. Methods. One eye from each of 20 adult, anesthetized brown Norway rats (group 1) was cannulated and simultaneously connected to a syringe and a pressure transducer with a chart recorder. We increased IOP from 15 to 45 mmHg in 5- mmHg increments and obtained 15 consecutive readings (ignoring instrument- generated averages) at each pressure increment with a Tono-Pen 2 tonometer. To test the tonopen's ability to measure unknown IOP, transducer pressures were varied randomly in 20 additional animals (group 2), and tonopen readings were obtained in masked fashion. Results. Plotting the mean tonopen readings for each animal against transducer IOP produced a regression formula of y = 4.54 + 0.79x (r = 0.98). Mean group 2 tonopen values plotted against transducer IOP yielded a regression formula of y = 4.75 + 0.78x (r = 0.94). A method comparison analysis showed that the tonopen significantly overestimates pressures at low IOP (≤ 15 mmHg), and it significantly underestimates pressures at high IOP (≥ 30 mmHg). Using two-way analysis of variance, it was determined that the group 2 data did not differ significantly from the group 1 data (P ≥ 0.76). Because of this consistency, we generated a correction factor with 95% prediction intervals for Tono-Pen readings. Conclusions. The Tono-Pen 2 can be used reliably to measure IOP in the normal rat eye.

AB - Purpose. The purpose of this study was to evaluate the Tono-Pen 2 tonometer for measuring intraocular pressure (IOP) in the living rat eye. Methods. One eye from each of 20 adult, anesthetized brown Norway rats (group 1) was cannulated and simultaneously connected to a syringe and a pressure transducer with a chart recorder. We increased IOP from 15 to 45 mmHg in 5- mmHg increments and obtained 15 consecutive readings (ignoring instrument- generated averages) at each pressure increment with a Tono-Pen 2 tonometer. To test the tonopen's ability to measure unknown IOP, transducer pressures were varied randomly in 20 additional animals (group 2), and tonopen readings were obtained in masked fashion. Results. Plotting the mean tonopen readings for each animal against transducer IOP produced a regression formula of y = 4.54 + 0.79x (r = 0.98). Mean group 2 tonopen values plotted against transducer IOP yielded a regression formula of y = 4.75 + 0.78x (r = 0.94). A method comparison analysis showed that the tonopen significantly overestimates pressures at low IOP (≤ 15 mmHg), and it significantly underestimates pressures at high IOP (≥ 30 mmHg). Using two-way analysis of variance, it was determined that the group 2 data did not differ significantly from the group 1 data (P ≥ 0.76). Because of this consistency, we generated a correction factor with 95% prediction intervals for Tono-Pen readings. Conclusions. The Tono-Pen 2 can be used reliably to measure IOP in the normal rat eye.

KW - intraocular pressure

KW - rat

KW - Tono-Pen

KW - tonometry

UR - http://www.scopus.com/inward/record.url?scp=0027457904&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0027457904&partnerID=8YFLogxK

M3 - Article

VL - 34

SP - 363

EP - 369

JO - Investigative Ophthalmology and Visual Science

JF - Investigative Ophthalmology and Visual Science

SN - 0146-0404

IS - 2

ER -