Intrauterine Vacuum-Induced Hemorrhage-Control Device for Rapid Treatment of Postpartum Hemorrhage

Mary E. D'Alton, Kara M. Rood, Marcela C. Smid, Hyagriv N. Simhan, Daniel W. Skupski, Akila Subramaniam, Kelly S. Gibson, Todd Rosen, Shannon M. Clark, Donald Dudley, Sara N. Iqbal, Michael J. Paglia, Christina M. Duzyj, Edward K. Chien, Karen J. Gibbins, Kathryn D. Wine, Nana Ama A. Bentum, Michelle A. Kominiarek, Methodius G. Tuuli, Dena Goffman

Research output: Contribution to journalArticlepeer-review

31 Scopus citations

Abstract

OBJECTIVE:To evaluate the effectiveness and safety of an intrauterine vacuum-induced hemorrhage-control device for postpartum hemorrhage treatment.METHODS:A multicenter, prospective, single-arm treatment study of a novel intrauterine device that uses low-level vacuum to induce uterine myometrial contraction to achieve control of abnormal postpartum uterine bleeding and postpartum hemorrhage was undertaken at 12 centers in the United States. The primary effectiveness endpoint was the proportion of participants in whom use of the intrauterine vacuum-induced hemorrhage-control device controlled abnormal bleeding without requiring escalating interventions. The primary safety endpoint was the incidence, severity, and seriousness of device-related adverse events. Secondary outcomes included time to bleeding control, rate of transfusion, and device usability scored by each investigator using the device.RESULTS:Of 107 participants enrolled with primary postpartum hemorrhage or abnormal postpartum uterine bleeding, 106 received any study treatment with the device connected to vacuum, and successful treatment was observed in 94% (100/106, 95% CI 88-98%) of these participants. In those 100 participants, definitive control of abnormal bleeding was reported in a median of 3 minutes (interquartile range 2.0-5.0) after connection to vacuum. Eight adverse events deemed possibly related to the device or procedure were reported, all of which were outlined as risks in the study and all of which resolved with treatment without serious clinical sequelae. Transfusion of 1-3 units of red blood cells was required in 35 participants, and five participants required 4 or more units of red blood cells. The majority of investigators reported the intrauterine vacuum-induced hemorrhage-control device as easy to use (98%) and would recommend it (97%).CONCLUSION:Intrauterine vacuum-induced hemorrhage control may provide a new rapid and effective treatment option for abnormal postpartum uterine bleeding or postpartum hemorrhage, with the potential to prevent severe maternal morbidity and mortality.FUNDING SOURCE:Alydia Health, Inc.CLINICAL TRIAL REGISTRATION:ClinicalTrials.gov, NCT02883673.

Original languageEnglish (US)
Pages (from-to)882-891
Number of pages10
JournalObstetrics and gynecology
Volume136
Issue number5
DOIs
StatePublished - Nov 1 2020

ASJC Scopus subject areas

  • Obstetrics and Gynecology

Fingerprint

Dive into the research topics of 'Intrauterine Vacuum-Induced Hemorrhage-Control Device for Rapid Treatment of Postpartum Hemorrhage'. Together they form a unique fingerprint.

Cite this