Importance of collimator selection for quantitative131I scintigraphy

H. D. Specht, Paul Brown, J. M. Hanada, A. A. Miley

Research output: Contribution to journalArticle

5 Citations (Scopus)

Abstract

Aberrant131I orthoiodohippurate renal plasma flow values were obtained when our image- based method was moved from an old to a new gamma camera. We suspected inadequate collimation as the reason for this problem because we observed increased septal penetration and scatter. Camera effects were largely eliminated because both our old and new cameras had 1/2 thick crystals. We report here the effects of two high and three medium energy collimators upon the fraction of syringe image count beyond the syringe image, and upon the correlation of these non-image count fractions with geometric collimator parameters. We also report the effects of these same collimators upon kidney-to-background count ratios, using patient relevant abdominal phantoms. The percentage counts beyond the syringe image varied from 33 to 57%. The phantom kidney-to-background ratios varied from 3.7 to 2.0. We conclude that image-based quantitative measurements are strongly dependent on collimator selection. The data also provide a user basis for collimator evaluation, selection and design, and have implications for131I and131In antibody imaging.

Original languageEnglish (US)
Pages (from-to)637-644
Number of pages8
JournalNuclear Medicine Communications
Volume12
Issue number7
StatePublished - 1991

Fingerprint

Syringes
Radionuclide Imaging
Iodohippuric Acid
Renal Plasma Flow
Kidney
Gamma Cameras
Antibodies

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging

Cite this

Specht, H. D., Brown, P., Hanada, J. M., & Miley, A. A. (1991). Importance of collimator selection for quantitative131I scintigraphy. Nuclear Medicine Communications, 12(7), 637-644.

Importance of collimator selection for quantitative131I scintigraphy. / Specht, H. D.; Brown, Paul; Hanada, J. M.; Miley, A. A.

In: Nuclear Medicine Communications, Vol. 12, No. 7, 1991, p. 637-644.

Research output: Contribution to journalArticle

Specht, HD, Brown, P, Hanada, JM & Miley, AA 1991, 'Importance of collimator selection for quantitative131I scintigraphy', Nuclear Medicine Communications, vol. 12, no. 7, pp. 637-644.
Specht, H. D. ; Brown, Paul ; Hanada, J. M. ; Miley, A. A. / Importance of collimator selection for quantitative131I scintigraphy. In: Nuclear Medicine Communications. 1991 ; Vol. 12, No. 7. pp. 637-644.
@article{0b5756043af94d2cac19054daf7e7bbd,
title = "Importance of collimator selection for quantitative131I scintigraphy",
abstract = "Aberrant131I orthoiodohippurate renal plasma flow values were obtained when our image- based method was moved from an old to a new gamma camera. We suspected inadequate collimation as the reason for this problem because we observed increased septal penetration and scatter. Camera effects were largely eliminated because both our old and new cameras had 1/2 thick crystals. We report here the effects of two high and three medium energy collimators upon the fraction of syringe image count beyond the syringe image, and upon the correlation of these non-image count fractions with geometric collimator parameters. We also report the effects of these same collimators upon kidney-to-background count ratios, using patient relevant abdominal phantoms. The percentage counts beyond the syringe image varied from 33 to 57{\%}. The phantom kidney-to-background ratios varied from 3.7 to 2.0. We conclude that image-based quantitative measurements are strongly dependent on collimator selection. The data also provide a user basis for collimator evaluation, selection and design, and have implications for131I and131In antibody imaging.",
author = "Specht, {H. D.} and Paul Brown and Hanada, {J. M.} and Miley, {A. A.}",
year = "1991",
language = "English (US)",
volume = "12",
pages = "637--644",
journal = "Nuclear Medicine Communications",
issn = "0143-3636",
publisher = "Lippincott Williams and Wilkins",
number = "7",

}

TY - JOUR

T1 - Importance of collimator selection for quantitative131I scintigraphy

AU - Specht, H. D.

AU - Brown, Paul

AU - Hanada, J. M.

AU - Miley, A. A.

PY - 1991

Y1 - 1991

N2 - Aberrant131I orthoiodohippurate renal plasma flow values were obtained when our image- based method was moved from an old to a new gamma camera. We suspected inadequate collimation as the reason for this problem because we observed increased septal penetration and scatter. Camera effects were largely eliminated because both our old and new cameras had 1/2 thick crystals. We report here the effects of two high and three medium energy collimators upon the fraction of syringe image count beyond the syringe image, and upon the correlation of these non-image count fractions with geometric collimator parameters. We also report the effects of these same collimators upon kidney-to-background count ratios, using patient relevant abdominal phantoms. The percentage counts beyond the syringe image varied from 33 to 57%. The phantom kidney-to-background ratios varied from 3.7 to 2.0. We conclude that image-based quantitative measurements are strongly dependent on collimator selection. The data also provide a user basis for collimator evaluation, selection and design, and have implications for131I and131In antibody imaging.

AB - Aberrant131I orthoiodohippurate renal plasma flow values were obtained when our image- based method was moved from an old to a new gamma camera. We suspected inadequate collimation as the reason for this problem because we observed increased septal penetration and scatter. Camera effects were largely eliminated because both our old and new cameras had 1/2 thick crystals. We report here the effects of two high and three medium energy collimators upon the fraction of syringe image count beyond the syringe image, and upon the correlation of these non-image count fractions with geometric collimator parameters. We also report the effects of these same collimators upon kidney-to-background count ratios, using patient relevant abdominal phantoms. The percentage counts beyond the syringe image varied from 33 to 57%. The phantom kidney-to-background ratios varied from 3.7 to 2.0. We conclude that image-based quantitative measurements are strongly dependent on collimator selection. The data also provide a user basis for collimator evaluation, selection and design, and have implications for131I and131In antibody imaging.

UR - http://www.scopus.com/inward/record.url?scp=84936906676&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84936906676&partnerID=8YFLogxK

M3 - Article

C2 - 1923156

AN - SCOPUS:84936906676

VL - 12

SP - 637

EP - 644

JO - Nuclear Medicine Communications

JF - Nuclear Medicine Communications

SN - 0143-3636

IS - 7

ER -