T1 - Determination of asymmetric cavity volumes using real-time three-dimensional echocardiography
T2 - An in vitro balloon model study
AU - Prakash, Kesavan
AU - Li, Xiaokui
AU - Hejmadi, Aarti
AU - Hashimoto, Ikuo
AU - Sahn, David J.
PY - 2004/4
Y1 - 2004/4
N2 - Objectives: We designed a new in vitro model to test the accuracy and reproducibility of real-time three-dimensional (RT3D) ultrasound imaging for determining a variety of asymmetric cavity volumes with aneurysm. Methods: Fifteen individual balloon models mimicking ventricular aneurysm were filled with water (170-322.5 ml) without air bubbles and kept in a compressor pump. Compression of the models produced only a change in shape of the balloon and no change in volume. The models were scanned with RT3D echocardiography (RT3DE) and the images recorded on an optical disk. Volumes were measured off line in two phases; maximal compression, where there was maximal change in shape and nil compression, where there was minimal or no change in shape. Volumes were measured by manual tracing technique of the inner border of B-scan images and compared with the drained volume of water from the balloon. Results: There was a high correlation between the drained volume and measured volume at maximal compression (equivalent to end-systole, r = 0.99, y = 0.99x + 3.69, SEE = 6.5 ml), between the drained volume and measured volume at nil compression (equivalent to end-diastole, r = 0.99, y = 0.94x + 12.07, SEE = 5.9 ml), and between volumes measured at maximal and nil compressions (r = 0.99, y = 0.94x + 10.55, SEE = 4.6 ml). Conclusion: The results of this experiment show that RT3DE can accurately measure the volumes of a variety of asymmetric ventricular cavities.
AB - Objectives: We designed a new in vitro model to test the accuracy and reproducibility of real-time three-dimensional (RT3D) ultrasound imaging for determining a variety of asymmetric cavity volumes with aneurysm. Methods: Fifteen individual balloon models mimicking ventricular aneurysm were filled with water (170-322.5 ml) without air bubbles and kept in a compressor pump. Compression of the models produced only a change in shape of the balloon and no change in volume. The models were scanned with RT3D echocardiography (RT3DE) and the images recorded on an optical disk. Volumes were measured off line in two phases; maximal compression, where there was maximal change in shape and nil compression, where there was minimal or no change in shape. Volumes were measured by manual tracing technique of the inner border of B-scan images and compared with the drained volume of water from the balloon. Results: There was a high correlation between the drained volume and measured volume at maximal compression (equivalent to end-systole, r = 0.99, y = 0.99x + 3.69, SEE = 6.5 ml), between the drained volume and measured volume at nil compression (equivalent to end-diastole, r = 0.99, y = 0.94x + 12.07, SEE = 5.9 ml), and between volumes measured at maximal and nil compressions (r = 0.99, y = 0.94x + 10.55, SEE = 4.6 ml). Conclusion: The results of this experiment show that RT3DE can accurately measure the volumes of a variety of asymmetric ventricular cavities.
KW - Real-time three-dimensional echocardiography
KW - Ventricular aneurysm
UR - http://www.scopus.com/inward/record.url?scp=1842778839&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=1842778839&partnerID=8YFLogxK
U2 - 10.1111/j.0742-2822.2004.03071.x
DO - 10.1111/j.0742-2822.2004.03071.x
M3 - Article
C2 - 15053788
AN - SCOPUS:1842778839
SN - 0742-2822
VL - 21
SP - 257
EP - 263
JO - Echocardiography
JF - Echocardiography
IS - 3
ER -