    Abstract We constructed an intracardiac vectorcardiogram from 3 configurations of intracardiac cardiovertor defibrilator (ICD) electrograms (EGMs). Six distinctive 3 lead combinations were selected out of five leads: can to right ventricular coil (RVC); RVC to superior vena cava coil (SVC); atrial lead tip (A-tip) to right ventricular (RV)-ring; can to RV-ring; RV-tip to RVC, in a patient with dual chamber ICD. Surface spatial QRS-T angle (119.8°) was similar to intracardiac spatial QRS-T angle derived from ICD EGMs combination A (101.3°), B (96.1°), C (92.8°), D (95.2), E (99.0), F (96.2) and median (101.5). Future validation of the novel method is needed.

    Original languageEnglish (US)
    Article number52059
    Pages (from-to)669-671
    Number of pages3
    JournalJournal of Electrocardiology
    Volume48
    Issue number4
    DOIs
    StatePublished - Jul 1 2015

    Keywords

    • Implantable cardioverter defibrillator
    • Intracardiac electrogram
    • Vectorcardiography

    ASJC Scopus subject areas

    • Cardiology and Cardiovascular Medicine

    Fingerprint Dive into the research topics of 'Construction of intracardiac vectorcardiogram from implantable cardioverter-defibrillator intracardiac electrograms'. Together they form a unique fingerprint.