Comparison of TG-43 dose calculations to pinpoint ion chamber and diamond detector measurements

Wolfram U. Laub

Research output: Contribution to journalReview articlepeer-review

4 Scopus citations

Abstract

Brachytherapy is an area of radiation therapy where the availability of high-resolution detectors with a low energy-dependence is of great importance. The suitability of two detectors is investigated in this study. Measurements in the proximity of an Ir-192 source were performed with a diamond detector type 60003 and a pinpoint chamber type 31006 (both PTW-Freiburg). For comparison, dose values were calculated with the dose calculation formalism recommended by the task group 43 (Nath et al 1995 Med. Phys. 22 209-34).

Original languageEnglish (US)
Pages (from-to)N315-N318
JournalPhysics in Medicine and Biology
Volume47
Issue number24
DOIs
StatePublished - Dec 21 2002
Externally publishedYes

ASJC Scopus subject areas

  • Radiological and Ultrasound Technology
  • Radiology Nuclear Medicine and imaging

Fingerprint

Dive into the research topics of 'Comparison of TG-43 dose calculations to pinpoint ion chamber and diamond detector measurements'. Together they form a unique fingerprint.