Clinical testing of the apnea prevention device: Proof of concept data

Research output: Contribution to journalArticle

2 Citations (Scopus)

Abstract

Introduction: Narcotic-induced respiratory depression is a major cause of perioperative morbidity and mortality. Current monitoring modalities are inadequate to detect and treat respiratory depression in postoperative patients. Intermittent nursing assessments, even if conducted frequently, may not capture the rapid onset of airway obstruction, apnea, and hypoxia that can occur in many of these patients. Continuous nursing observation, as in an intensive care unit setting, is cost-prohibitive and impractical, given the large number of patients at risk. In an effort to address this problem, the author has created and tested the Apnea Prevention Device (APD), which is designed to detect the onset of hypoxia and instantly intervene to restore respiration and oxygen saturation in narcotized patients. Methods: The prototype APD used in this study consisted of a laptop computer running custom software, a pulse oximeter, and a nerve stimulator. Oxygen saturation data were acquired by the computer from the pulse oximeter, and stimuli to the patient were delivered either by headphones (verbal prompts) or a nerve stimulator (cutaneous). The APD program was written to analyze oximetry data and when indicated, deliver a series of stimuli of increasing intensity to arouse patients from narcosis. The device was tested on surgical patients in the postanesthesia care unit. An intervention delivered by the APD was scored as a success if the patient took a large tidal volume breath (as evidenced by chest rise) and there was a subsequent increase in oxygen saturation. The APD maintained a data log of oxygen saturations and interventions. In a subset of patients, it was possible to compare the functioning of the APD with routine nursing care. Results: A total of 125 interventions were delivered by the APD to 10 patients with a 97% success rate. The depth of desaturations was less when the APD was in use than when patients received routine 1:1 nursing care. When the APD was functioning, the frequency with which nurses prompted the subjects to breathe was dramatically reduced. Discussion: This study demonstrates that the prototype APD can successfully treat narcotic-induced respiratory depression in postoperative patients and does so in a manner that is superior to that provided by routine 1:1 nursing care. Such a device has the potential to decrease the morbidity associated with narcotic-induced respiratory depression in postoperative patients.

Original languageEnglish (US)
Pages (from-to)582-586
Number of pages5
JournalAnesthesia and Analgesia
Volume112
Issue number3
DOIs
StatePublished - Mar 2011

Fingerprint

Apnea
Equipment and Supplies
Respiratory Insufficiency
Narcotics
Nursing Care
Oxygen
Pulse
Nursing Assessment
Morbidity
Stupor
Oximetry
Tidal Volume
Airway Obstruction
Intensive Care Units
Respiration
Nursing
Thorax
Software
Nurses
Observation

ASJC Scopus subject areas

  • Anesthesiology and Pain Medicine

Cite this

Clinical testing of the apnea prevention device : Proof of concept data. / Zornow, Mark.

In: Anesthesia and Analgesia, Vol. 112, No. 3, 03.2011, p. 582-586.

Research output: Contribution to journalArticle

@article{913ad772c07748929f414d6a915d9750,
title = "Clinical testing of the apnea prevention device: Proof of concept data",
abstract = "Introduction: Narcotic-induced respiratory depression is a major cause of perioperative morbidity and mortality. Current monitoring modalities are inadequate to detect and treat respiratory depression in postoperative patients. Intermittent nursing assessments, even if conducted frequently, may not capture the rapid onset of airway obstruction, apnea, and hypoxia that can occur in many of these patients. Continuous nursing observation, as in an intensive care unit setting, is cost-prohibitive and impractical, given the large number of patients at risk. In an effort to address this problem, the author has created and tested the Apnea Prevention Device (APD), which is designed to detect the onset of hypoxia and instantly intervene to restore respiration and oxygen saturation in narcotized patients. Methods: The prototype APD used in this study consisted of a laptop computer running custom software, a pulse oximeter, and a nerve stimulator. Oxygen saturation data were acquired by the computer from the pulse oximeter, and stimuli to the patient were delivered either by headphones (verbal prompts) or a nerve stimulator (cutaneous). The APD program was written to analyze oximetry data and when indicated, deliver a series of stimuli of increasing intensity to arouse patients from narcosis. The device was tested on surgical patients in the postanesthesia care unit. An intervention delivered by the APD was scored as a success if the patient took a large tidal volume breath (as evidenced by chest rise) and there was a subsequent increase in oxygen saturation. The APD maintained a data log of oxygen saturations and interventions. In a subset of patients, it was possible to compare the functioning of the APD with routine nursing care. Results: A total of 125 interventions were delivered by the APD to 10 patients with a 97{\%} success rate. The depth of desaturations was less when the APD was in use than when patients received routine 1:1 nursing care. When the APD was functioning, the frequency with which nurses prompted the subjects to breathe was dramatically reduced. Discussion: This study demonstrates that the prototype APD can successfully treat narcotic-induced respiratory depression in postoperative patients and does so in a manner that is superior to that provided by routine 1:1 nursing care. Such a device has the potential to decrease the morbidity associated with narcotic-induced respiratory depression in postoperative patients.",
author = "Mark Zornow",
year = "2011",
month = "3",
doi = "10.1213/ANE.0b013e318204e3cb",
language = "English (US)",
volume = "112",
pages = "582--586",
journal = "Anesthesia and Analgesia",
issn = "0003-2999",
publisher = "Lippincott Williams and Wilkins",
number = "3",

}

TY - JOUR

T1 - Clinical testing of the apnea prevention device

T2 - Proof of concept data

AU - Zornow, Mark

PY - 2011/3

Y1 - 2011/3

N2 - Introduction: Narcotic-induced respiratory depression is a major cause of perioperative morbidity and mortality. Current monitoring modalities are inadequate to detect and treat respiratory depression in postoperative patients. Intermittent nursing assessments, even if conducted frequently, may not capture the rapid onset of airway obstruction, apnea, and hypoxia that can occur in many of these patients. Continuous nursing observation, as in an intensive care unit setting, is cost-prohibitive and impractical, given the large number of patients at risk. In an effort to address this problem, the author has created and tested the Apnea Prevention Device (APD), which is designed to detect the onset of hypoxia and instantly intervene to restore respiration and oxygen saturation in narcotized patients. Methods: The prototype APD used in this study consisted of a laptop computer running custom software, a pulse oximeter, and a nerve stimulator. Oxygen saturation data were acquired by the computer from the pulse oximeter, and stimuli to the patient were delivered either by headphones (verbal prompts) or a nerve stimulator (cutaneous). The APD program was written to analyze oximetry data and when indicated, deliver a series of stimuli of increasing intensity to arouse patients from narcosis. The device was tested on surgical patients in the postanesthesia care unit. An intervention delivered by the APD was scored as a success if the patient took a large tidal volume breath (as evidenced by chest rise) and there was a subsequent increase in oxygen saturation. The APD maintained a data log of oxygen saturations and interventions. In a subset of patients, it was possible to compare the functioning of the APD with routine nursing care. Results: A total of 125 interventions were delivered by the APD to 10 patients with a 97% success rate. The depth of desaturations was less when the APD was in use than when patients received routine 1:1 nursing care. When the APD was functioning, the frequency with which nurses prompted the subjects to breathe was dramatically reduced. Discussion: This study demonstrates that the prototype APD can successfully treat narcotic-induced respiratory depression in postoperative patients and does so in a manner that is superior to that provided by routine 1:1 nursing care. Such a device has the potential to decrease the morbidity associated with narcotic-induced respiratory depression in postoperative patients.

AB - Introduction: Narcotic-induced respiratory depression is a major cause of perioperative morbidity and mortality. Current monitoring modalities are inadequate to detect and treat respiratory depression in postoperative patients. Intermittent nursing assessments, even if conducted frequently, may not capture the rapid onset of airway obstruction, apnea, and hypoxia that can occur in many of these patients. Continuous nursing observation, as in an intensive care unit setting, is cost-prohibitive and impractical, given the large number of patients at risk. In an effort to address this problem, the author has created and tested the Apnea Prevention Device (APD), which is designed to detect the onset of hypoxia and instantly intervene to restore respiration and oxygen saturation in narcotized patients. Methods: The prototype APD used in this study consisted of a laptop computer running custom software, a pulse oximeter, and a nerve stimulator. Oxygen saturation data were acquired by the computer from the pulse oximeter, and stimuli to the patient were delivered either by headphones (verbal prompts) or a nerve stimulator (cutaneous). The APD program was written to analyze oximetry data and when indicated, deliver a series of stimuli of increasing intensity to arouse patients from narcosis. The device was tested on surgical patients in the postanesthesia care unit. An intervention delivered by the APD was scored as a success if the patient took a large tidal volume breath (as evidenced by chest rise) and there was a subsequent increase in oxygen saturation. The APD maintained a data log of oxygen saturations and interventions. In a subset of patients, it was possible to compare the functioning of the APD with routine nursing care. Results: A total of 125 interventions were delivered by the APD to 10 patients with a 97% success rate. The depth of desaturations was less when the APD was in use than when patients received routine 1:1 nursing care. When the APD was functioning, the frequency with which nurses prompted the subjects to breathe was dramatically reduced. Discussion: This study demonstrates that the prototype APD can successfully treat narcotic-induced respiratory depression in postoperative patients and does so in a manner that is superior to that provided by routine 1:1 nursing care. Such a device has the potential to decrease the morbidity associated with narcotic-induced respiratory depression in postoperative patients.

UR - http://www.scopus.com/inward/record.url?scp=79952442986&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=79952442986&partnerID=8YFLogxK

U2 - 10.1213/ANE.0b013e318204e3cb

DO - 10.1213/ANE.0b013e318204e3cb

M3 - Article

C2 - 21233491

AN - SCOPUS:79952442986

VL - 112

SP - 582

EP - 586

JO - Anesthesia and Analgesia

JF - Anesthesia and Analgesia

SN - 0003-2999

IS - 3

ER -