Biomechanical comparison of a novel multilevel hex-head pedicle screw design with a conventional head design

Qi Liu, Donald R. Olson, Frederick W. Tiley, Marie Shea, Matt Smits, Robert Hart

Research output: Contribution to journalArticle

1 Citation (Scopus)

Abstract

The objective of the study was to determine the biomechanical effect during insertion of multilevel hex-head design pedicle screws compared to a conventional screw-head design. Eighteen lumbar vertebrae and thoracic vertebrae from human cadavers were instrumented with a novel, multilevel hexagonal head pedicle screw on one side and a conventional head pedicle screw on the contralateral side. Screws were inserted at a constant rate and insertion and removal torques were recorded. A further 14 lumbar and thoracic vertebrae were used to test alterability of screw direction and operational effort required. Electromagnetic sensors recorded the change in angular direction for both screw and screwdriver. The force applied through the insertion screwdriver required to produce the directional change was also recorded. No significant differences were found between the two screw types for insertion or removal torque in either lumbar or thoracic vertebrae. Multilevel hex-head screws had significantly greater directional alterability than conventional head screws in both lumbar and thoracic specimens. Multilevel hex-head screws also required less force applied through the screwdriver than conventional screws to alter direction of screw insertion in both lumbar and thoracic specimens. The multilevel hex-head design did not affect the insertion or removal torque in comparison to a conventional head design.

Original languageEnglish (US)
Pages (from-to)1115-1120
Number of pages6
JournalJournal of Orthopaedic Research
Volume25
Issue number8
DOIs
StatePublished - Aug 2007

Fingerprint

Head
Thoracic Vertebrae
Lumbar Vertebrae
Torque
Thorax
Pedicle Screws
Electromagnetic Phenomena
Cadaver
Direction compound

Keywords

  • Biomechanical comparison
  • Conventional head design
  • Hex-head screws

ASJC Scopus subject areas

  • Orthopedics and Sports Medicine

Cite this

Biomechanical comparison of a novel multilevel hex-head pedicle screw design with a conventional head design. / Liu, Qi; Olson, Donald R.; Tiley, Frederick W.; Shea, Marie; Smits, Matt; Hart, Robert.

In: Journal of Orthopaedic Research, Vol. 25, No. 8, 08.2007, p. 1115-1120.

Research output: Contribution to journalArticle

Liu, Qi ; Olson, Donald R. ; Tiley, Frederick W. ; Shea, Marie ; Smits, Matt ; Hart, Robert. / Biomechanical comparison of a novel multilevel hex-head pedicle screw design with a conventional head design. In: Journal of Orthopaedic Research. 2007 ; Vol. 25, No. 8. pp. 1115-1120.
@article{302da85f535148be9c3131612d0aa066,
title = "Biomechanical comparison of a novel multilevel hex-head pedicle screw design with a conventional head design",
abstract = "The objective of the study was to determine the biomechanical effect during insertion of multilevel hex-head design pedicle screws compared to a conventional screw-head design. Eighteen lumbar vertebrae and thoracic vertebrae from human cadavers were instrumented with a novel, multilevel hexagonal head pedicle screw on one side and a conventional head pedicle screw on the contralateral side. Screws were inserted at a constant rate and insertion and removal torques were recorded. A further 14 lumbar and thoracic vertebrae were used to test alterability of screw direction and operational effort required. Electromagnetic sensors recorded the change in angular direction for both screw and screwdriver. The force applied through the insertion screwdriver required to produce the directional change was also recorded. No significant differences were found between the two screw types for insertion or removal torque in either lumbar or thoracic vertebrae. Multilevel hex-head screws had significantly greater directional alterability than conventional head screws in both lumbar and thoracic specimens. Multilevel hex-head screws also required less force applied through the screwdriver than conventional screws to alter direction of screw insertion in both lumbar and thoracic specimens. The multilevel hex-head design did not affect the insertion or removal torque in comparison to a conventional head design.",
keywords = "Biomechanical comparison, Conventional head design, Hex-head screws",
author = "Qi Liu and Olson, {Donald R.} and Tiley, {Frederick W.} and Marie Shea and Matt Smits and Robert Hart",
year = "2007",
month = "8",
doi = "10.1002/jor.20401",
language = "English (US)",
volume = "25",
pages = "1115--1120",
journal = "Journal of Orthopaedic Research",
issn = "0736-0266",
publisher = "John Wiley and Sons Inc.",
number = "8",

}

TY - JOUR

T1 - Biomechanical comparison of a novel multilevel hex-head pedicle screw design with a conventional head design

AU - Liu, Qi

AU - Olson, Donald R.

AU - Tiley, Frederick W.

AU - Shea, Marie

AU - Smits, Matt

AU - Hart, Robert

PY - 2007/8

Y1 - 2007/8

N2 - The objective of the study was to determine the biomechanical effect during insertion of multilevel hex-head design pedicle screws compared to a conventional screw-head design. Eighteen lumbar vertebrae and thoracic vertebrae from human cadavers were instrumented with a novel, multilevel hexagonal head pedicle screw on one side and a conventional head pedicle screw on the contralateral side. Screws were inserted at a constant rate and insertion and removal torques were recorded. A further 14 lumbar and thoracic vertebrae were used to test alterability of screw direction and operational effort required. Electromagnetic sensors recorded the change in angular direction for both screw and screwdriver. The force applied through the insertion screwdriver required to produce the directional change was also recorded. No significant differences were found between the two screw types for insertion or removal torque in either lumbar or thoracic vertebrae. Multilevel hex-head screws had significantly greater directional alterability than conventional head screws in both lumbar and thoracic specimens. Multilevel hex-head screws also required less force applied through the screwdriver than conventional screws to alter direction of screw insertion in both lumbar and thoracic specimens. The multilevel hex-head design did not affect the insertion or removal torque in comparison to a conventional head design.

AB - The objective of the study was to determine the biomechanical effect during insertion of multilevel hex-head design pedicle screws compared to a conventional screw-head design. Eighteen lumbar vertebrae and thoracic vertebrae from human cadavers were instrumented with a novel, multilevel hexagonal head pedicle screw on one side and a conventional head pedicle screw on the contralateral side. Screws were inserted at a constant rate and insertion and removal torques were recorded. A further 14 lumbar and thoracic vertebrae were used to test alterability of screw direction and operational effort required. Electromagnetic sensors recorded the change in angular direction for both screw and screwdriver. The force applied through the insertion screwdriver required to produce the directional change was also recorded. No significant differences were found between the two screw types for insertion or removal torque in either lumbar or thoracic vertebrae. Multilevel hex-head screws had significantly greater directional alterability than conventional head screws in both lumbar and thoracic specimens. Multilevel hex-head screws also required less force applied through the screwdriver than conventional screws to alter direction of screw insertion in both lumbar and thoracic specimens. The multilevel hex-head design did not affect the insertion or removal torque in comparison to a conventional head design.

KW - Biomechanical comparison

KW - Conventional head design

KW - Hex-head screws

UR - http://www.scopus.com/inward/record.url?scp=34547734920&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=34547734920&partnerID=8YFLogxK

U2 - 10.1002/jor.20401

DO - 10.1002/jor.20401

M3 - Article

C2 - 17444509

AN - SCOPUS:34547734920

VL - 25

SP - 1115

EP - 1120

JO - Journal of Orthopaedic Research

JF - Journal of Orthopaedic Research

SN - 0736-0266

IS - 8

ER -