We tested the accuracy and reproducibility of knowledge-based reconstruction (KBR) for measuring right ventricular (RV) volume and function. KBR enables rapid assessment of the right ventricle from sparse user input by referencing a database. KBR generates a 3-dimensional surface to fit points that the user enters at anatomic landmarks. We measured the RV volume using KBR from magnetic resonance images in 20 patients with repaired tetralogy of Fallot at end-diastole and end-systole. We entered points in the long- and short-axis and/or oblique views. The true volume was computed by manually tracing the RV borders for 3-dimensional reconstruction using the piecewise smooth subdivision surface method. The reference database included 54 patients with tetralogy of Fallot patients. The KBR values agreed closely with the true values for the end-diastolic volume (r = 0.993), end-systolic volume (r = 0.992), and ejection fraction (EF; r = 0.930). KBR slightly overestimated the end-diastolic volume (4 ± 10 ml, p = NS), end-systolic volume (1 ± 9 ml, p = NS), and EF (4 ± 3%, p = NS). No bias in the error was found by Bland-Altman analysis (p = NS for end-diastolic and end-systolic volume and EF). The KBR volumes had approached the true volumes (235 ± 93 vs 243 ± 93, p = 0.012, r = 0.978 for end-diastolic and end-systolic volumes combined) already after the first run and the entry of 19 ± 3 points. In conclusion, KBR provided accurate measurement of the RV volume and EF with minimal user input. KBR is a clinically feasible alternative to full manual tracing of the heart borders from imaging data.

Original languageEnglish (US)
Pages (from-to)993-999
Number of pages7
JournalAmerican Journal of Cardiology
Volume105
Issue number7
DOIs
StatePublished - Apr 1 2010

Fingerprint

Right Ventricular Function
Tetralogy of Fallot
montirelin
Databases
Anatomic Landmarks
Diastole
Systole
Heart Ventricles
Magnetic Resonance Spectroscopy

ASJC Scopus subject areas

  • Cardiology and Cardiovascular Medicine

Cite this

Accuracy of Knowledge-Based Reconstruction for Measurement of Right Ventricular Volume and Function in Patients With Tetralogy of Fallot. / Sheehan, Florence H.; Kilner, Philip J.; Sahn, David; Vick, G. Wesley; Stout, Karen K.; Ge, Shuping; Helbing, Willem A.; Lewin, Mark; Shurman, Alan J.; Buechel, Emanuela Valsangiacomo; Litt, Harold I.; Waiss, Mary Pierre.

In: American Journal of Cardiology, Vol. 105, No. 7, 01.04.2010, p. 993-999.

Research output: Contribution to journalArticle

Sheehan, FH, Kilner, PJ, Sahn, D, Vick, GW, Stout, KK, Ge, S, Helbing, WA, Lewin, M, Shurman, AJ, Buechel, EV, Litt, HI & Waiss, MP 2010, 'Accuracy of Knowledge-Based Reconstruction for Measurement of Right Ventricular Volume and Function in Patients With Tetralogy of Fallot', American Journal of Cardiology, vol. 105, no. 7, pp. 993-999. https://doi.org/10.1016/j.amjcard.2009.11.032
Sheehan, Florence H. ; Kilner, Philip J. ; Sahn, David ; Vick, G. Wesley ; Stout, Karen K. ; Ge, Shuping ; Helbing, Willem A. ; Lewin, Mark ; Shurman, Alan J. ; Buechel, Emanuela Valsangiacomo ; Litt, Harold I. ; Waiss, Mary Pierre. / Accuracy of Knowledge-Based Reconstruction for Measurement of Right Ventricular Volume and Function in Patients With Tetralogy of Fallot. In: American Journal of Cardiology. 2010 ; Vol. 105, No. 7. pp. 993-999.
@article{b3e5878408ae4666b7a0153fe4acb3b3,
title = "Accuracy of Knowledge-Based Reconstruction for Measurement of Right Ventricular Volume and Function in Patients With Tetralogy of Fallot",
abstract = "We tested the accuracy and reproducibility of knowledge-based reconstruction (KBR) for measuring right ventricular (RV) volume and function. KBR enables rapid assessment of the right ventricle from sparse user input by referencing a database. KBR generates a 3-dimensional surface to fit points that the user enters at anatomic landmarks. We measured the RV volume using KBR from magnetic resonance images in 20 patients with repaired tetralogy of Fallot at end-diastole and end-systole. We entered points in the long- and short-axis and/or oblique views. The true volume was computed by manually tracing the RV borders for 3-dimensional reconstruction using the piecewise smooth subdivision surface method. The reference database included 54 patients with tetralogy of Fallot patients. The KBR values agreed closely with the true values for the end-diastolic volume (r = 0.993), end-systolic volume (r = 0.992), and ejection fraction (EF; r = 0.930). KBR slightly overestimated the end-diastolic volume (4 ± 10 ml, p = NS), end-systolic volume (1 ± 9 ml, p = NS), and EF (4 ± 3{\%}, p = NS). No bias in the error was found by Bland-Altman analysis (p = NS for end-diastolic and end-systolic volume and EF). The KBR volumes had approached the true volumes (235 ± 93 vs 243 ± 93, p = 0.012, r = 0.978 for end-diastolic and end-systolic volumes combined) already after the first run and the entry of 19 ± 3 points. In conclusion, KBR provided accurate measurement of the RV volume and EF with minimal user input. KBR is a clinically feasible alternative to full manual tracing of the heart borders from imaging data.",
author = "Sheehan, {Florence H.} and Kilner, {Philip J.} and David Sahn and Vick, {G. Wesley} and Stout, {Karen K.} and Shuping Ge and Helbing, {Willem A.} and Mark Lewin and Shurman, {Alan J.} and Buechel, {Emanuela Valsangiacomo} and Litt, {Harold I.} and Waiss, {Mary Pierre}",
year = "2010",
month = "4",
day = "1",
doi = "10.1016/j.amjcard.2009.11.032",
language = "English (US)",
volume = "105",
pages = "993--999",
journal = "American Journal of Cardiology",
issn = "0002-9149",
publisher = "Elsevier Inc.",
number = "7",

}

TY - JOUR

T1 - Accuracy of Knowledge-Based Reconstruction for Measurement of Right Ventricular Volume and Function in Patients With Tetralogy of Fallot

AU - Sheehan, Florence H.

AU - Kilner, Philip J.

AU - Sahn, David

AU - Vick, G. Wesley

AU - Stout, Karen K.

AU - Ge, Shuping

AU - Helbing, Willem A.

AU - Lewin, Mark

AU - Shurman, Alan J.

AU - Buechel, Emanuela Valsangiacomo

AU - Litt, Harold I.

AU - Waiss, Mary Pierre

PY - 2010/4/1

Y1 - 2010/4/1

N2 - We tested the accuracy and reproducibility of knowledge-based reconstruction (KBR) for measuring right ventricular (RV) volume and function. KBR enables rapid assessment of the right ventricle from sparse user input by referencing a database. KBR generates a 3-dimensional surface to fit points that the user enters at anatomic landmarks. We measured the RV volume using KBR from magnetic resonance images in 20 patients with repaired tetralogy of Fallot at end-diastole and end-systole. We entered points in the long- and short-axis and/or oblique views. The true volume was computed by manually tracing the RV borders for 3-dimensional reconstruction using the piecewise smooth subdivision surface method. The reference database included 54 patients with tetralogy of Fallot patients. The KBR values agreed closely with the true values for the end-diastolic volume (r = 0.993), end-systolic volume (r = 0.992), and ejection fraction (EF; r = 0.930). KBR slightly overestimated the end-diastolic volume (4 ± 10 ml, p = NS), end-systolic volume (1 ± 9 ml, p = NS), and EF (4 ± 3%, p = NS). No bias in the error was found by Bland-Altman analysis (p = NS for end-diastolic and end-systolic volume and EF). The KBR volumes had approached the true volumes (235 ± 93 vs 243 ± 93, p = 0.012, r = 0.978 for end-diastolic and end-systolic volumes combined) already after the first run and the entry of 19 ± 3 points. In conclusion, KBR provided accurate measurement of the RV volume and EF with minimal user input. KBR is a clinically feasible alternative to full manual tracing of the heart borders from imaging data.

AB - We tested the accuracy and reproducibility of knowledge-based reconstruction (KBR) for measuring right ventricular (RV) volume and function. KBR enables rapid assessment of the right ventricle from sparse user input by referencing a database. KBR generates a 3-dimensional surface to fit points that the user enters at anatomic landmarks. We measured the RV volume using KBR from magnetic resonance images in 20 patients with repaired tetralogy of Fallot at end-diastole and end-systole. We entered points in the long- and short-axis and/or oblique views. The true volume was computed by manually tracing the RV borders for 3-dimensional reconstruction using the piecewise smooth subdivision surface method. The reference database included 54 patients with tetralogy of Fallot patients. The KBR values agreed closely with the true values for the end-diastolic volume (r = 0.993), end-systolic volume (r = 0.992), and ejection fraction (EF; r = 0.930). KBR slightly overestimated the end-diastolic volume (4 ± 10 ml, p = NS), end-systolic volume (1 ± 9 ml, p = NS), and EF (4 ± 3%, p = NS). No bias in the error was found by Bland-Altman analysis (p = NS for end-diastolic and end-systolic volume and EF). The KBR volumes had approached the true volumes (235 ± 93 vs 243 ± 93, p = 0.012, r = 0.978 for end-diastolic and end-systolic volumes combined) already after the first run and the entry of 19 ± 3 points. In conclusion, KBR provided accurate measurement of the RV volume and EF with minimal user input. KBR is a clinically feasible alternative to full manual tracing of the heart borders from imaging data.

UR - http://www.scopus.com/inward/record.url?scp=77949567270&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=77949567270&partnerID=8YFLogxK

U2 - 10.1016/j.amjcard.2009.11.032

DO - 10.1016/j.amjcard.2009.11.032

M3 - Article

VL - 105

SP - 993

EP - 999

JO - American Journal of Cardiology

JF - American Journal of Cardiology

SN - 0002-9149

IS - 7

ER -